Data and methods to estimate fetal dose from fluoroscopically guided prophylactic hypogastric artery balloon occlusion

Purpose: To provide data for estimation of fetal radiation dose (DF ) from prophylactic hypogastric artery balloon occlusion (HABO) procedures. Methods: The Monte-Carlo-N-particle (MCNP) transport code and mathematical phantoms representing a pregnant patient at the ninth month of gestation were emp...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical physics (Lancaster) 2016-06, Vol.43 (6), p.2990-2997
Hauptverfasser: Solomou, G., Perisinakis, K., Tsetis, D., Stratakis, J., Damilakis, J.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose: To provide data for estimation of fetal radiation dose (DF ) from prophylactic hypogastric artery balloon occlusion (HABO) procedures. Methods: The Monte-Carlo-N-particle (MCNP) transport code and mathematical phantoms representing a pregnant patient at the ninth month of gestation were employed. PA, RAO 20° and LAO 20° fluoroscopy projections of left and right internal iliac arteries were simulated. Projection-specific normalized fetal dose (NFD) data were produced for various beam qualities. The effects of projection angle, x-ray field location relative to the fetus, field size, maternal body size, and fetal size on NFD were investigated. Presented NFD values were compared to corresponding values derived using a physical anthropomorphic phantom simulating pregnancy at the third trimester and thermoluminescence dosimeters. Results: NFD did not considerably vary when projection angle was altered by ±5°, whereas it was found to markedly depend on tube voltage, filtration, x-ray field location and size, and maternal body size. Differences in NFD < 7.5% were observed for naturally expected variations in fetal size. A difference of less than 13.5% was observed between NFD values estimated by MCNP and direct measurements. Conclusions: Data and methods provided allow for reliable estimation of radiation burden to the fetus from HABO.
ISSN:0094-2405
2473-4209
DOI:10.1118/1.4950718