Quality assurance and independent dosimetry for an intraoperative x-ray device
Purpose: Quality assurance is an essential component of accurate and safe radiotherapy delivery, and should include measurements which are independent of manufacturer-provided calibration. However, the physical and dosimetric properties of the INTRABEAM compact mobile 50 kV x-ray source are differen...
Gespeichert in:
Veröffentlicht in: | Medical Physics 2012-11, Vol.39 (11), p.6908-6920 |
---|---|
1. Verfasser: | |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose:
Quality assurance is an essential component of accurate and safe radiotherapy delivery, and should include measurements which are independent of manufacturer-provided calibration. However, the physical and dosimetric properties of the INTRABEAM compact mobile 50 kV x-ray source are different from conventional kilovoltage therapy units and few reports describe methods for independent checks, frequencies, or tolerances for quality assurance tests.
Methods:
Based on the available evidence and local experience, methods are described for determination of the key dosimetric parameters: beam quality, output, isotropy, and depth doses. Internal system checks are also described, along with measurements of long-term stability.
Results:
A small volume parallel plate ionization chamber in a liquid water tank is the gold standard for measurements with this unit, but solid water-equivalent materials, thermoluminescent dosimeters and radiochromic film can all be used as practical alternatives with an accuracy of 5%–10%. The main cause of measurement uncertainty is positioning of the detector in the steep dose gradient, but energy dependence should also be considered.
Conclusions:
A quality assurance schedule with suggested tolerances is proposed, which includes both internal tests, before each treatment and on a monthly basis, and independent tests every year or after servicing or recalibration. |
---|---|
ISSN: | 0094-2405 2473-4209 |