Performance measurements for the PET/CT Discovery-600 using NEMA NU 2-2007 standards

Purpose: The aim of this study was to assess the performance measurements of the new PET/CT system Discovery-600 (D-600, GEMS, Milwaukee, WI). Methods: Performance measures were obtained with the National Electrical Manufacturers Association (NEMA) NU 2-2007 procedures. Results: The transverse (axia...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical physics (Lancaster) 2011-02, Vol.38 (2), p.968-974
Hauptverfasser: De Ponti, E., Morzenti, S., Guerra, L., Pasquali, C., Arosio, M., Bettinardi, V., Crespi, A., Gilardi, M. C., Messa, C.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 974
container_issue 2
container_start_page 968
container_title Medical physics (Lancaster)
container_volume 38
creator De Ponti, E.
Morzenti, S.
Guerra, L.
Pasquali, C.
Arosio, M.
Bettinardi, V.
Crespi, A.
Gilardi, M. C.
Messa, C.
description Purpose: The aim of this study was to assess the performance measurements of the new PET/CT system Discovery-600 (D-600, GEMS, Milwaukee, WI). Methods: Performance measures were obtained with the National Electrical Manufacturers Association (NEMA) NU 2-2007 procedures. Results: The transverse (axial) spatial resolution FWHMs were 4.9 (5.6) mm and 5.6 (6.4) mm at 1 and 10 cm off axis, respectively. The sensitivity (average at 0 and 10 cm) was 9.6 cps/kBq. The scatter fraction was 36.6% (low energy threshold: 425 keV). The NEC peak rate ( k = 1 ) was 75.2 kcps at 12.9 kBq/cc. The hot contrasts for 10, 13, 17, and 22 mm spheres were 41%, 51%, 62%, and 73% and the cold contrasts for 28 and 37 mm spheres were 68% and 72%. Conclusions: The Discovery-600 has good performance for the NEMA NU 2-2007 parameters, particularly in improved sensitivity compared to the scanners of the same Discovery family, D-ST and D-STE.
doi_str_mv 10.1118/1.3544655
format Article
fullrecord <record><control><sourceid>proquest_wiley</sourceid><recordid>TN_cdi_wiley_primary_10_1118_1_3544655_MP4655</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>859759473</sourcerecordid><originalsourceid>FETCH-LOGICAL-c4865-5d2c4b31da8cf4ab8fea30b6bbfff866a029389dcb1b4e404ed5e064082d76723</originalsourceid><addsrcrecordid>eNp90F1rFDEUBuAgil2rF_4BCXhRFKY9-ZyZy7KuVujHXmyvQyY5sSM7kzWZqey_d8qspSD26kB48nLOS8h7BqeMseqMnQolpVbqBVlwWYpCcqhfkgVALQsuQR2RNzn_BAAtFLwmR5xJxUshFmSzxhRi6mzvkHZo85iww37IdHqlwx3S9WpzttzQL2128R7TvtAAdMxt_4Ner67O6fUt5QUHKGkebO9t8vkteRXsNuO7wzwmt19Xm-VFcXnz7fvy_LJwstKqUJ472QjmbeWCtE0V0ApodNOEECqtLfBaVLV3DWskSpDoFYKWUHFf6pKLY_Jxzo15aE127YDuzsW-RzcYPnWga6YmdTKrXYq_RsyD6aZbcLu1PcYxm0rVpaqn2ib5aZYuxZwTBrNLbWfT3jAwD00bZg5NT_bDIXVsOvSP8m-1Eyhm8Lvd4v7_SeZqfQj8PPuHO-zQxv7xz31MT_zOh-fwv6v-Ad7ZnrY</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>859759473</pqid></control><display><type>article</type><title>Performance measurements for the PET/CT Discovery-600 using NEMA NU 2-2007 standards</title><source>MEDLINE</source><source>Wiley Online Library Journals Frontfile Complete</source><source>Alma/SFX Local Collection</source><creator>De Ponti, E. ; Morzenti, S. ; Guerra, L. ; Pasquali, C. ; Arosio, M. ; Bettinardi, V. ; Crespi, A. ; Gilardi, M. C. ; Messa, C.</creator><creatorcontrib>De Ponti, E. ; Morzenti, S. ; Guerra, L. ; Pasquali, C. ; Arosio, M. ; Bettinardi, V. ; Crespi, A. ; Gilardi, M. C. ; Messa, C.</creatorcontrib><description>Purpose: The aim of this study was to assess the performance measurements of the new PET/CT system Discovery-600 (D-600, GEMS, Milwaukee, WI). Methods: Performance measures were obtained with the National Electrical Manufacturers Association (NEMA) NU 2-2007 procedures. Results: The transverse (axial) spatial resolution FWHMs were 4.9 (5.6) mm and 5.6 (6.4) mm at 1 and 10 cm off axis, respectively. The sensitivity (average at 0 and 10 cm) was 9.6 cps/kBq. The scatter fraction was 36.6% (low energy threshold: 425 keV). The NEC peak rate ( k = 1 ) was 75.2 kcps at 12.9 kBq/cc. The hot contrasts for 10, 13, 17, and 22 mm spheres were 41%, 51%, 62%, and 73% and the cold contrasts for 28 and 37 mm spheres were 68% and 72%. Conclusions: The Discovery-600 has good performance for the NEMA NU 2-2007 parameters, particularly in improved sensitivity compared to the scanners of the same Discovery family, D-ST and D-STE.</description><identifier>ISSN: 0094-2405</identifier><identifier>EISSN: 2473-4209</identifier><identifier>DOI: 10.1118/1.3544655</identifier><identifier>PMID: 21452733</identifier><identifier>CODEN: MPHYA6</identifier><language>eng</language><publisher>United States: American Association of Physicists in Medicine</publisher><subject>60 APPLIED LIFE SCIENCES ; benchmark testing ; BENCHMARKS ; biomedical equipment ; Biomedical instrumentation and transducers, including micro‐electro‐mechanical systems (MEMS) ; Computed tomography ; computerised tomography ; Discovery‐600 ; Electricity ; Image reconstruction ; Image scanners ; Image sensors ; INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY ; KEV RANGE 100-1000 ; Lungs ; measurement standards ; Medical image quality ; Medical image reconstruction ; Medical imaging ; NEMA NU 2‐2007 ; PET/CT ; POSITRON COMPUTED TOMOGRAPHY ; positron emission tomography ; Positron emission tomography (PET) ; Positron-Emission Tomography - standards ; Reference Standards ; SENSITIVITY ; SPATIAL RESOLUTION ; Standards and calibration ; TESTING ; Tomography, X-Ray Computed - standards ; United States</subject><ispartof>Medical physics (Lancaster), 2011-02, Vol.38 (2), p.968-974</ispartof><rights>American Association of Physicists in Medicine</rights><rights>2011 American Association of Physicists in Medicine</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c4865-5d2c4b31da8cf4ab8fea30b6bbfff866a029389dcb1b4e404ed5e064082d76723</citedby><cites>FETCH-LOGICAL-c4865-5d2c4b31da8cf4ab8fea30b6bbfff866a029389dcb1b4e404ed5e064082d76723</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://onlinelibrary.wiley.com/doi/pdf/10.1118%2F1.3544655$$EPDF$$P50$$Gwiley$$H</linktopdf><linktohtml>$$Uhttps://onlinelibrary.wiley.com/doi/full/10.1118%2F1.3544655$$EHTML$$P50$$Gwiley$$H</linktohtml><link.rule.ids>230,314,776,780,881,1411,27901,27902,45550,45551</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/21452733$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink><backlink>$$Uhttps://www.osti.gov/biblio/22096915$$D View this record in Osti.gov$$Hfree_for_read</backlink></links><search><creatorcontrib>De Ponti, E.</creatorcontrib><creatorcontrib>Morzenti, S.</creatorcontrib><creatorcontrib>Guerra, L.</creatorcontrib><creatorcontrib>Pasquali, C.</creatorcontrib><creatorcontrib>Arosio, M.</creatorcontrib><creatorcontrib>Bettinardi, V.</creatorcontrib><creatorcontrib>Crespi, A.</creatorcontrib><creatorcontrib>Gilardi, M. C.</creatorcontrib><creatorcontrib>Messa, C.</creatorcontrib><title>Performance measurements for the PET/CT Discovery-600 using NEMA NU 2-2007 standards</title><title>Medical physics (Lancaster)</title><addtitle>Med Phys</addtitle><description>Purpose: The aim of this study was to assess the performance measurements of the new PET/CT system Discovery-600 (D-600, GEMS, Milwaukee, WI). Methods: Performance measures were obtained with the National Electrical Manufacturers Association (NEMA) NU 2-2007 procedures. Results: The transverse (axial) spatial resolution FWHMs were 4.9 (5.6) mm and 5.6 (6.4) mm at 1 and 10 cm off axis, respectively. The sensitivity (average at 0 and 10 cm) was 9.6 cps/kBq. The scatter fraction was 36.6% (low energy threshold: 425 keV). The NEC peak rate ( k = 1 ) was 75.2 kcps at 12.9 kBq/cc. The hot contrasts for 10, 13, 17, and 22 mm spheres were 41%, 51%, 62%, and 73% and the cold contrasts for 28 and 37 mm spheres were 68% and 72%. Conclusions: The Discovery-600 has good performance for the NEMA NU 2-2007 parameters, particularly in improved sensitivity compared to the scanners of the same Discovery family, D-ST and D-STE.</description><subject>60 APPLIED LIFE SCIENCES</subject><subject>benchmark testing</subject><subject>BENCHMARKS</subject><subject>biomedical equipment</subject><subject>Biomedical instrumentation and transducers, including micro‐electro‐mechanical systems (MEMS)</subject><subject>Computed tomography</subject><subject>computerised tomography</subject><subject>Discovery‐600</subject><subject>Electricity</subject><subject>Image reconstruction</subject><subject>Image scanners</subject><subject>Image sensors</subject><subject>INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY</subject><subject>KEV RANGE 100-1000</subject><subject>Lungs</subject><subject>measurement standards</subject><subject>Medical image quality</subject><subject>Medical image reconstruction</subject><subject>Medical imaging</subject><subject>NEMA NU 2‐2007</subject><subject>PET/CT</subject><subject>POSITRON COMPUTED TOMOGRAPHY</subject><subject>positron emission tomography</subject><subject>Positron emission tomography (PET)</subject><subject>Positron-Emission Tomography - standards</subject><subject>Reference Standards</subject><subject>SENSITIVITY</subject><subject>SPATIAL RESOLUTION</subject><subject>Standards and calibration</subject><subject>TESTING</subject><subject>Tomography, X-Ray Computed - standards</subject><subject>United States</subject><issn>0094-2405</issn><issn>2473-4209</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2011</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNp90F1rFDEUBuAgil2rF_4BCXhRFKY9-ZyZy7KuVujHXmyvQyY5sSM7kzWZqey_d8qspSD26kB48nLOS8h7BqeMseqMnQolpVbqBVlwWYpCcqhfkgVALQsuQR2RNzn_BAAtFLwmR5xJxUshFmSzxhRi6mzvkHZo85iww37IdHqlwx3S9WpzttzQL2128R7TvtAAdMxt_4Ner67O6fUt5QUHKGkebO9t8vkteRXsNuO7wzwmt19Xm-VFcXnz7fvy_LJwstKqUJ472QjmbeWCtE0V0ApodNOEECqtLfBaVLV3DWskSpDoFYKWUHFf6pKLY_Jxzo15aE127YDuzsW-RzcYPnWga6YmdTKrXYq_RsyD6aZbcLu1PcYxm0rVpaqn2ib5aZYuxZwTBrNLbWfT3jAwD00bZg5NT_bDIXVsOvSP8m-1Eyhm8Lvd4v7_SeZqfQj8PPuHO-zQxv7xz31MT_zOh-fwv6v-Ad7ZnrY</recordid><startdate>201102</startdate><enddate>201102</enddate><creator>De Ponti, E.</creator><creator>Morzenti, S.</creator><creator>Guerra, L.</creator><creator>Pasquali, C.</creator><creator>Arosio, M.</creator><creator>Bettinardi, V.</creator><creator>Crespi, A.</creator><creator>Gilardi, M. C.</creator><creator>Messa, C.</creator><general>American Association of Physicists in Medicine</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><scope>OTOTI</scope></search><sort><creationdate>201102</creationdate><title>Performance measurements for the PET/CT Discovery-600 using NEMA NU 2-2007 standards</title><author>De Ponti, E. ; Morzenti, S. ; Guerra, L. ; Pasquali, C. ; Arosio, M. ; Bettinardi, V. ; Crespi, A. ; Gilardi, M. C. ; Messa, C.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c4865-5d2c4b31da8cf4ab8fea30b6bbfff866a029389dcb1b4e404ed5e064082d76723</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2011</creationdate><topic>60 APPLIED LIFE SCIENCES</topic><topic>benchmark testing</topic><topic>BENCHMARKS</topic><topic>biomedical equipment</topic><topic>Biomedical instrumentation and transducers, including micro‐electro‐mechanical systems (MEMS)</topic><topic>Computed tomography</topic><topic>computerised tomography</topic><topic>Discovery‐600</topic><topic>Electricity</topic><topic>Image reconstruction</topic><topic>Image scanners</topic><topic>Image sensors</topic><topic>INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY</topic><topic>KEV RANGE 100-1000</topic><topic>Lungs</topic><topic>measurement standards</topic><topic>Medical image quality</topic><topic>Medical image reconstruction</topic><topic>Medical imaging</topic><topic>NEMA NU 2‐2007</topic><topic>PET/CT</topic><topic>POSITRON COMPUTED TOMOGRAPHY</topic><topic>positron emission tomography</topic><topic>Positron emission tomography (PET)</topic><topic>Positron-Emission Tomography - standards</topic><topic>Reference Standards</topic><topic>SENSITIVITY</topic><topic>SPATIAL RESOLUTION</topic><topic>Standards and calibration</topic><topic>TESTING</topic><topic>Tomography, X-Ray Computed - standards</topic><topic>United States</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>De Ponti, E.</creatorcontrib><creatorcontrib>Morzenti, S.</creatorcontrib><creatorcontrib>Guerra, L.</creatorcontrib><creatorcontrib>Pasquali, C.</creatorcontrib><creatorcontrib>Arosio, M.</creatorcontrib><creatorcontrib>Bettinardi, V.</creatorcontrib><creatorcontrib>Crespi, A.</creatorcontrib><creatorcontrib>Gilardi, M. C.</creatorcontrib><creatorcontrib>Messa, C.</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><collection>OSTI.GOV</collection><jtitle>Medical physics (Lancaster)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>De Ponti, E.</au><au>Morzenti, S.</au><au>Guerra, L.</au><au>Pasquali, C.</au><au>Arosio, M.</au><au>Bettinardi, V.</au><au>Crespi, A.</au><au>Gilardi, M. C.</au><au>Messa, C.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Performance measurements for the PET/CT Discovery-600 using NEMA NU 2-2007 standards</atitle><jtitle>Medical physics (Lancaster)</jtitle><addtitle>Med Phys</addtitle><date>2011-02</date><risdate>2011</risdate><volume>38</volume><issue>2</issue><spage>968</spage><epage>974</epage><pages>968-974</pages><issn>0094-2405</issn><eissn>2473-4209</eissn><coden>MPHYA6</coden><abstract>Purpose: The aim of this study was to assess the performance measurements of the new PET/CT system Discovery-600 (D-600, GEMS, Milwaukee, WI). Methods: Performance measures were obtained with the National Electrical Manufacturers Association (NEMA) NU 2-2007 procedures. Results: The transverse (axial) spatial resolution FWHMs were 4.9 (5.6) mm and 5.6 (6.4) mm at 1 and 10 cm off axis, respectively. The sensitivity (average at 0 and 10 cm) was 9.6 cps/kBq. The scatter fraction was 36.6% (low energy threshold: 425 keV). The NEC peak rate ( k = 1 ) was 75.2 kcps at 12.9 kBq/cc. The hot contrasts for 10, 13, 17, and 22 mm spheres were 41%, 51%, 62%, and 73% and the cold contrasts for 28 and 37 mm spheres were 68% and 72%. Conclusions: The Discovery-600 has good performance for the NEMA NU 2-2007 parameters, particularly in improved sensitivity compared to the scanners of the same Discovery family, D-ST and D-STE.</abstract><cop>United States</cop><pub>American Association of Physicists in Medicine</pub><pmid>21452733</pmid><doi>10.1118/1.3544655</doi><tpages>7</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0094-2405
ispartof Medical physics (Lancaster), 2011-02, Vol.38 (2), p.968-974
issn 0094-2405
2473-4209
language eng
recordid cdi_wiley_primary_10_1118_1_3544655_MP4655
source MEDLINE; Wiley Online Library Journals Frontfile Complete; Alma/SFX Local Collection
subjects 60 APPLIED LIFE SCIENCES
benchmark testing
BENCHMARKS
biomedical equipment
Biomedical instrumentation and transducers, including micro‐electro‐mechanical systems (MEMS)
Computed tomography
computerised tomography
Discovery‐600
Electricity
Image reconstruction
Image scanners
Image sensors
INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
KEV RANGE 100-1000
Lungs
measurement standards
Medical image quality
Medical image reconstruction
Medical imaging
NEMA NU 2‐2007
PET/CT
POSITRON COMPUTED TOMOGRAPHY
positron emission tomography
Positron emission tomography (PET)
Positron-Emission Tomography - standards
Reference Standards
SENSITIVITY
SPATIAL RESOLUTION
Standards and calibration
TESTING
Tomography, X-Ray Computed - standards
United States
title Performance measurements for the PET/CT Discovery-600 using NEMA NU 2-2007 standards
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-05T23%3A09%3A46IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_wiley&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Performance%20measurements%20for%20the%20PET/CT%20Discovery-600%20using%20NEMA%20NU%202-2007%20standards&rft.jtitle=Medical%20physics%20(Lancaster)&rft.au=De%20Ponti,%20E.&rft.date=2011-02&rft.volume=38&rft.issue=2&rft.spage=968&rft.epage=974&rft.pages=968-974&rft.issn=0094-2405&rft.eissn=2473-4209&rft.coden=MPHYA6&rft_id=info:doi/10.1118/1.3544655&rft_dat=%3Cproquest_wiley%3E859759473%3C/proquest_wiley%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=859759473&rft_id=info:pmid/21452733&rfr_iscdi=true