Performance measurements for the PET/CT Discovery-600 using NEMA NU 2-2007 standards
Purpose: The aim of this study was to assess the performance measurements of the new PET/CT system Discovery-600 (D-600, GEMS, Milwaukee, WI). Methods: Performance measures were obtained with the National Electrical Manufacturers Association (NEMA) NU 2-2007 procedures. Results: The transverse (axia...
Gespeichert in:
Veröffentlicht in: | Medical physics (Lancaster) 2011-02, Vol.38 (2), p.968-974 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 974 |
---|---|
container_issue | 2 |
container_start_page | 968 |
container_title | Medical physics (Lancaster) |
container_volume | 38 |
creator | De Ponti, E. Morzenti, S. Guerra, L. Pasquali, C. Arosio, M. Bettinardi, V. Crespi, A. Gilardi, M. C. Messa, C. |
description | Purpose:
The aim of this study was to assess the performance measurements of the new PET/CT system Discovery-600 (D-600, GEMS, Milwaukee, WI).
Methods:
Performance measures were obtained with the National Electrical Manufacturers Association (NEMA) NU 2-2007 procedures.
Results:
The transverse (axial) spatial resolution FWHMs were 4.9 (5.6) mm and 5.6 (6.4) mm at 1 and 10 cm off axis, respectively. The sensitivity (average at 0 and 10 cm) was 9.6 cps/kBq. The scatter fraction was 36.6% (low energy threshold: 425 keV). The NEC peak rate
(
k
=
1
)
was 75.2 kcps at 12.9 kBq/cc. The hot contrasts for 10, 13, 17, and 22 mm spheres were 41%, 51%, 62%, and 73% and the cold contrasts for 28 and 37 mm spheres were 68% and 72%.
Conclusions:
The Discovery-600 has good performance for the NEMA NU 2-2007 parameters, particularly in improved sensitivity compared to the scanners of the same Discovery family, D-ST and D-STE. |
doi_str_mv | 10.1118/1.3544655 |
format | Article |
fullrecord | <record><control><sourceid>proquest_wiley</sourceid><recordid>TN_cdi_wiley_primary_10_1118_1_3544655_MP4655</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>859759473</sourcerecordid><originalsourceid>FETCH-LOGICAL-c4865-5d2c4b31da8cf4ab8fea30b6bbfff866a029389dcb1b4e404ed5e064082d76723</originalsourceid><addsrcrecordid>eNp90F1rFDEUBuAgil2rF_4BCXhRFKY9-ZyZy7KuVujHXmyvQyY5sSM7kzWZqey_d8qspSD26kB48nLOS8h7BqeMseqMnQolpVbqBVlwWYpCcqhfkgVALQsuQR2RNzn_BAAtFLwmR5xJxUshFmSzxhRi6mzvkHZo85iww37IdHqlwx3S9WpzttzQL2128R7TvtAAdMxt_4Ner67O6fUt5QUHKGkebO9t8vkteRXsNuO7wzwmt19Xm-VFcXnz7fvy_LJwstKqUJ472QjmbeWCtE0V0ApodNOEECqtLfBaVLV3DWskSpDoFYKWUHFf6pKLY_Jxzo15aE127YDuzsW-RzcYPnWga6YmdTKrXYq_RsyD6aZbcLu1PcYxm0rVpaqn2ib5aZYuxZwTBrNLbWfT3jAwD00bZg5NT_bDIXVsOvSP8m-1Eyhm8Lvd4v7_SeZqfQj8PPuHO-zQxv7xz31MT_zOh-fwv6v-Ad7ZnrY</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>859759473</pqid></control><display><type>article</type><title>Performance measurements for the PET/CT Discovery-600 using NEMA NU 2-2007 standards</title><source>MEDLINE</source><source>Wiley Online Library Journals Frontfile Complete</source><source>Alma/SFX Local Collection</source><creator>De Ponti, E. ; Morzenti, S. ; Guerra, L. ; Pasquali, C. ; Arosio, M. ; Bettinardi, V. ; Crespi, A. ; Gilardi, M. C. ; Messa, C.</creator><creatorcontrib>De Ponti, E. ; Morzenti, S. ; Guerra, L. ; Pasquali, C. ; Arosio, M. ; Bettinardi, V. ; Crespi, A. ; Gilardi, M. C. ; Messa, C.</creatorcontrib><description>Purpose:
The aim of this study was to assess the performance measurements of the new PET/CT system Discovery-600 (D-600, GEMS, Milwaukee, WI).
Methods:
Performance measures were obtained with the National Electrical Manufacturers Association (NEMA) NU 2-2007 procedures.
Results:
The transverse (axial) spatial resolution FWHMs were 4.9 (5.6) mm and 5.6 (6.4) mm at 1 and 10 cm off axis, respectively. The sensitivity (average at 0 and 10 cm) was 9.6 cps/kBq. The scatter fraction was 36.6% (low energy threshold: 425 keV). The NEC peak rate
(
k
=
1
)
was 75.2 kcps at 12.9 kBq/cc. The hot contrasts for 10, 13, 17, and 22 mm spheres were 41%, 51%, 62%, and 73% and the cold contrasts for 28 and 37 mm spheres were 68% and 72%.
Conclusions:
The Discovery-600 has good performance for the NEMA NU 2-2007 parameters, particularly in improved sensitivity compared to the scanners of the same Discovery family, D-ST and D-STE.</description><identifier>ISSN: 0094-2405</identifier><identifier>EISSN: 2473-4209</identifier><identifier>DOI: 10.1118/1.3544655</identifier><identifier>PMID: 21452733</identifier><identifier>CODEN: MPHYA6</identifier><language>eng</language><publisher>United States: American Association of Physicists in Medicine</publisher><subject>60 APPLIED LIFE SCIENCES ; benchmark testing ; BENCHMARKS ; biomedical equipment ; Biomedical instrumentation and transducers, including micro‐electro‐mechanical systems (MEMS) ; Computed tomography ; computerised tomography ; Discovery‐600 ; Electricity ; Image reconstruction ; Image scanners ; Image sensors ; INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY ; KEV RANGE 100-1000 ; Lungs ; measurement standards ; Medical image quality ; Medical image reconstruction ; Medical imaging ; NEMA NU 2‐2007 ; PET/CT ; POSITRON COMPUTED TOMOGRAPHY ; positron emission tomography ; Positron emission tomography (PET) ; Positron-Emission Tomography - standards ; Reference Standards ; SENSITIVITY ; SPATIAL RESOLUTION ; Standards and calibration ; TESTING ; Tomography, X-Ray Computed - standards ; United States</subject><ispartof>Medical physics (Lancaster), 2011-02, Vol.38 (2), p.968-974</ispartof><rights>American Association of Physicists in Medicine</rights><rights>2011 American Association of Physicists in Medicine</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c4865-5d2c4b31da8cf4ab8fea30b6bbfff866a029389dcb1b4e404ed5e064082d76723</citedby><cites>FETCH-LOGICAL-c4865-5d2c4b31da8cf4ab8fea30b6bbfff866a029389dcb1b4e404ed5e064082d76723</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://onlinelibrary.wiley.com/doi/pdf/10.1118%2F1.3544655$$EPDF$$P50$$Gwiley$$H</linktopdf><linktohtml>$$Uhttps://onlinelibrary.wiley.com/doi/full/10.1118%2F1.3544655$$EHTML$$P50$$Gwiley$$H</linktohtml><link.rule.ids>230,314,776,780,881,1411,27901,27902,45550,45551</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/21452733$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink><backlink>$$Uhttps://www.osti.gov/biblio/22096915$$D View this record in Osti.gov$$Hfree_for_read</backlink></links><search><creatorcontrib>De Ponti, E.</creatorcontrib><creatorcontrib>Morzenti, S.</creatorcontrib><creatorcontrib>Guerra, L.</creatorcontrib><creatorcontrib>Pasquali, C.</creatorcontrib><creatorcontrib>Arosio, M.</creatorcontrib><creatorcontrib>Bettinardi, V.</creatorcontrib><creatorcontrib>Crespi, A.</creatorcontrib><creatorcontrib>Gilardi, M. C.</creatorcontrib><creatorcontrib>Messa, C.</creatorcontrib><title>Performance measurements for the PET/CT Discovery-600 using NEMA NU 2-2007 standards</title><title>Medical physics (Lancaster)</title><addtitle>Med Phys</addtitle><description>Purpose:
The aim of this study was to assess the performance measurements of the new PET/CT system Discovery-600 (D-600, GEMS, Milwaukee, WI).
Methods:
Performance measures were obtained with the National Electrical Manufacturers Association (NEMA) NU 2-2007 procedures.
Results:
The transverse (axial) spatial resolution FWHMs were 4.9 (5.6) mm and 5.6 (6.4) mm at 1 and 10 cm off axis, respectively. The sensitivity (average at 0 and 10 cm) was 9.6 cps/kBq. The scatter fraction was 36.6% (low energy threshold: 425 keV). The NEC peak rate
(
k
=
1
)
was 75.2 kcps at 12.9 kBq/cc. The hot contrasts for 10, 13, 17, and 22 mm spheres were 41%, 51%, 62%, and 73% and the cold contrasts for 28 and 37 mm spheres were 68% and 72%.
Conclusions:
The Discovery-600 has good performance for the NEMA NU 2-2007 parameters, particularly in improved sensitivity compared to the scanners of the same Discovery family, D-ST and D-STE.</description><subject>60 APPLIED LIFE SCIENCES</subject><subject>benchmark testing</subject><subject>BENCHMARKS</subject><subject>biomedical equipment</subject><subject>Biomedical instrumentation and transducers, including micro‐electro‐mechanical systems (MEMS)</subject><subject>Computed tomography</subject><subject>computerised tomography</subject><subject>Discovery‐600</subject><subject>Electricity</subject><subject>Image reconstruction</subject><subject>Image scanners</subject><subject>Image sensors</subject><subject>INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY</subject><subject>KEV RANGE 100-1000</subject><subject>Lungs</subject><subject>measurement standards</subject><subject>Medical image quality</subject><subject>Medical image reconstruction</subject><subject>Medical imaging</subject><subject>NEMA NU 2‐2007</subject><subject>PET/CT</subject><subject>POSITRON COMPUTED TOMOGRAPHY</subject><subject>positron emission tomography</subject><subject>Positron emission tomography (PET)</subject><subject>Positron-Emission Tomography - standards</subject><subject>Reference Standards</subject><subject>SENSITIVITY</subject><subject>SPATIAL RESOLUTION</subject><subject>Standards and calibration</subject><subject>TESTING</subject><subject>Tomography, X-Ray Computed - standards</subject><subject>United States</subject><issn>0094-2405</issn><issn>2473-4209</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2011</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNp90F1rFDEUBuAgil2rF_4BCXhRFKY9-ZyZy7KuVujHXmyvQyY5sSM7kzWZqey_d8qspSD26kB48nLOS8h7BqeMseqMnQolpVbqBVlwWYpCcqhfkgVALQsuQR2RNzn_BAAtFLwmR5xJxUshFmSzxhRi6mzvkHZo85iww37IdHqlwx3S9WpzttzQL2128R7TvtAAdMxt_4Ner67O6fUt5QUHKGkebO9t8vkteRXsNuO7wzwmt19Xm-VFcXnz7fvy_LJwstKqUJ472QjmbeWCtE0V0ApodNOEECqtLfBaVLV3DWskSpDoFYKWUHFf6pKLY_Jxzo15aE127YDuzsW-RzcYPnWga6YmdTKrXYq_RsyD6aZbcLu1PcYxm0rVpaqn2ib5aZYuxZwTBrNLbWfT3jAwD00bZg5NT_bDIXVsOvSP8m-1Eyhm8Lvd4v7_SeZqfQj8PPuHO-zQxv7xz31MT_zOh-fwv6v-Ad7ZnrY</recordid><startdate>201102</startdate><enddate>201102</enddate><creator>De Ponti, E.</creator><creator>Morzenti, S.</creator><creator>Guerra, L.</creator><creator>Pasquali, C.</creator><creator>Arosio, M.</creator><creator>Bettinardi, V.</creator><creator>Crespi, A.</creator><creator>Gilardi, M. C.</creator><creator>Messa, C.</creator><general>American Association of Physicists in Medicine</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><scope>OTOTI</scope></search><sort><creationdate>201102</creationdate><title>Performance measurements for the PET/CT Discovery-600 using NEMA NU 2-2007 standards</title><author>De Ponti, E. ; Morzenti, S. ; Guerra, L. ; Pasquali, C. ; Arosio, M. ; Bettinardi, V. ; Crespi, A. ; Gilardi, M. C. ; Messa, C.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c4865-5d2c4b31da8cf4ab8fea30b6bbfff866a029389dcb1b4e404ed5e064082d76723</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2011</creationdate><topic>60 APPLIED LIFE SCIENCES</topic><topic>benchmark testing</topic><topic>BENCHMARKS</topic><topic>biomedical equipment</topic><topic>Biomedical instrumentation and transducers, including micro‐electro‐mechanical systems (MEMS)</topic><topic>Computed tomography</topic><topic>computerised tomography</topic><topic>Discovery‐600</topic><topic>Electricity</topic><topic>Image reconstruction</topic><topic>Image scanners</topic><topic>Image sensors</topic><topic>INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY</topic><topic>KEV RANGE 100-1000</topic><topic>Lungs</topic><topic>measurement standards</topic><topic>Medical image quality</topic><topic>Medical image reconstruction</topic><topic>Medical imaging</topic><topic>NEMA NU 2‐2007</topic><topic>PET/CT</topic><topic>POSITRON COMPUTED TOMOGRAPHY</topic><topic>positron emission tomography</topic><topic>Positron emission tomography (PET)</topic><topic>Positron-Emission Tomography - standards</topic><topic>Reference Standards</topic><topic>SENSITIVITY</topic><topic>SPATIAL RESOLUTION</topic><topic>Standards and calibration</topic><topic>TESTING</topic><topic>Tomography, X-Ray Computed - standards</topic><topic>United States</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>De Ponti, E.</creatorcontrib><creatorcontrib>Morzenti, S.</creatorcontrib><creatorcontrib>Guerra, L.</creatorcontrib><creatorcontrib>Pasquali, C.</creatorcontrib><creatorcontrib>Arosio, M.</creatorcontrib><creatorcontrib>Bettinardi, V.</creatorcontrib><creatorcontrib>Crespi, A.</creatorcontrib><creatorcontrib>Gilardi, M. C.</creatorcontrib><creatorcontrib>Messa, C.</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><collection>OSTI.GOV</collection><jtitle>Medical physics (Lancaster)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>De Ponti, E.</au><au>Morzenti, S.</au><au>Guerra, L.</au><au>Pasquali, C.</au><au>Arosio, M.</au><au>Bettinardi, V.</au><au>Crespi, A.</au><au>Gilardi, M. C.</au><au>Messa, C.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Performance measurements for the PET/CT Discovery-600 using NEMA NU 2-2007 standards</atitle><jtitle>Medical physics (Lancaster)</jtitle><addtitle>Med Phys</addtitle><date>2011-02</date><risdate>2011</risdate><volume>38</volume><issue>2</issue><spage>968</spage><epage>974</epage><pages>968-974</pages><issn>0094-2405</issn><eissn>2473-4209</eissn><coden>MPHYA6</coden><abstract>Purpose:
The aim of this study was to assess the performance measurements of the new PET/CT system Discovery-600 (D-600, GEMS, Milwaukee, WI).
Methods:
Performance measures were obtained with the National Electrical Manufacturers Association (NEMA) NU 2-2007 procedures.
Results:
The transverse (axial) spatial resolution FWHMs were 4.9 (5.6) mm and 5.6 (6.4) mm at 1 and 10 cm off axis, respectively. The sensitivity (average at 0 and 10 cm) was 9.6 cps/kBq. The scatter fraction was 36.6% (low energy threshold: 425 keV). The NEC peak rate
(
k
=
1
)
was 75.2 kcps at 12.9 kBq/cc. The hot contrasts for 10, 13, 17, and 22 mm spheres were 41%, 51%, 62%, and 73% and the cold contrasts for 28 and 37 mm spheres were 68% and 72%.
Conclusions:
The Discovery-600 has good performance for the NEMA NU 2-2007 parameters, particularly in improved sensitivity compared to the scanners of the same Discovery family, D-ST and D-STE.</abstract><cop>United States</cop><pub>American Association of Physicists in Medicine</pub><pmid>21452733</pmid><doi>10.1118/1.3544655</doi><tpages>7</tpages></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0094-2405 |
ispartof | Medical physics (Lancaster), 2011-02, Vol.38 (2), p.968-974 |
issn | 0094-2405 2473-4209 |
language | eng |
recordid | cdi_wiley_primary_10_1118_1_3544655_MP4655 |
source | MEDLINE; Wiley Online Library Journals Frontfile Complete; Alma/SFX Local Collection |
subjects | 60 APPLIED LIFE SCIENCES benchmark testing BENCHMARKS biomedical equipment Biomedical instrumentation and transducers, including micro‐electro‐mechanical systems (MEMS) Computed tomography computerised tomography Discovery‐600 Electricity Image reconstruction Image scanners Image sensors INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY KEV RANGE 100-1000 Lungs measurement standards Medical image quality Medical image reconstruction Medical imaging NEMA NU 2‐2007 PET/CT POSITRON COMPUTED TOMOGRAPHY positron emission tomography Positron emission tomography (PET) Positron-Emission Tomography - standards Reference Standards SENSITIVITY SPATIAL RESOLUTION Standards and calibration TESTING Tomography, X-Ray Computed - standards United States |
title | Performance measurements for the PET/CT Discovery-600 using NEMA NU 2-2007 standards |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-05T23%3A09%3A46IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_wiley&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Performance%20measurements%20for%20the%20PET/CT%20Discovery-600%20using%20NEMA%20NU%202-2007%20standards&rft.jtitle=Medical%20physics%20(Lancaster)&rft.au=De%20Ponti,%20E.&rft.date=2011-02&rft.volume=38&rft.issue=2&rft.spage=968&rft.epage=974&rft.pages=968-974&rft.issn=0094-2405&rft.eissn=2473-4209&rft.coden=MPHYA6&rft_id=info:doi/10.1118/1.3544655&rft_dat=%3Cproquest_wiley%3E859759473%3C/proquest_wiley%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=859759473&rft_id=info:pmid/21452733&rfr_iscdi=true |