Performance measurements for the PET/CT Discovery-600 using NEMA NU 2-2007 standards
Purpose: The aim of this study was to assess the performance measurements of the new PET/CT system Discovery-600 (D-600, GEMS, Milwaukee, WI). Methods: Performance measures were obtained with the National Electrical Manufacturers Association (NEMA) NU 2-2007 procedures. Results: The transverse (axia...
Gespeichert in:
Veröffentlicht in: | Medical physics (Lancaster) 2011-02, Vol.38 (2), p.968-974 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose:
The aim of this study was to assess the performance measurements of the new PET/CT system Discovery-600 (D-600, GEMS, Milwaukee, WI).
Methods:
Performance measures were obtained with the National Electrical Manufacturers Association (NEMA) NU 2-2007 procedures.
Results:
The transverse (axial) spatial resolution FWHMs were 4.9 (5.6) mm and 5.6 (6.4) mm at 1 and 10 cm off axis, respectively. The sensitivity (average at 0 and 10 cm) was 9.6 cps/kBq. The scatter fraction was 36.6% (low energy threshold: 425 keV). The NEC peak rate
(
k
=
1
)
was 75.2 kcps at 12.9 kBq/cc. The hot contrasts for 10, 13, 17, and 22 mm spheres were 41%, 51%, 62%, and 73% and the cold contrasts for 28 and 37 mm spheres were 68% and 72%.
Conclusions:
The Discovery-600 has good performance for the NEMA NU 2-2007 parameters, particularly in improved sensitivity compared to the scanners of the same Discovery family, D-ST and D-STE. |
---|---|