Monte Carlo techniques for scattering foil design and dosimetry in total skin electron irradiations
Total skin electron irradiation (TSEI) with single fields requires large electron beams having good dose uniformity, d max at the skin surface, and low bremsstrahlung contamination. To satisfy these requirements, energy degraders and scattering foils have to be specially designed for the given accel...
Gespeichert in:
Veröffentlicht in: | Medical physics (Lancaster) 2005-06, Vol.32 (6), p.1460-1468 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Total skin electron irradiation (TSEI) with single fields requires large electron beams having good dose uniformity,
d
max
at the skin surface, and low bremsstrahlung contamination. To satisfy these requirements, energy degraders and scattering foils have to be specially designed for the given accelerator and treatment room. We used Monte Carlo (MC) techniques based on EGS4 user codes (BEAM, DOSXYZ, and DOSRZ) as a guide in the beam modifier design of our TSEI system. The dosimetric characteristics at the treatment distance of
382
cm
source-to-surface distance (SSD) were verified experimentally using a linear array of 47 ion chambers, a parallel plate chamber, and radiochromic film. By matching MC simulations to standard beam measurements at
100
cm
SSD, the parameters of the electron beam incident on the vacuum window were determined. Best match was achieved assuming that electrons were monoenergetic at
6.72
MeV
, parallel, and distributed in a circular pattern having a Gaussian radial distribution with full width at half
maximum
=
0.13
cm
. These parameters were then used to simulate our TSEI unit with various scattering foils. Two of the foils were fabricated and experimentally evaluated by measuring off-axis dose uniformity and depth doses. A scattering foil, consisting of a
12
×
12
cm
2
aluminum plate of
0.6
cm
thickness and placed at isocenter perpendicular to the beam direction, was considered optimal. It produced a beam that was flat within
±
3
%
up to
60
cm
off-axis distance, dropped by not more than 8% at a distance of
90
cm
, and had an x-ray contamination of
<
3
%
. For stationary beams, MC-computed
d
max
,
R
p
, and
R
50
agreed with measurements within
0.5
mm
. The MC-predicted surface dose of the rotating phantom was 41% of the dose rate at
d
max
of the stationary phantom, whereas our calculations based on a semiempirical formula in the literature yielded a drop to 42%. The MC simulations provided the guideline of beam modifier design for TSEI and estimated the dosimetric performance for stationary and rotational irradiations. |
---|---|
ISSN: | 0094-2405 2473-4209 |
DOI: | 10.1118/1.1924368 |