Visible-light optical coherence tomography platform for the characterization of the skin barrier

We demonstrate a free-space, trolley-mountable Fourier domain visible-light optical coherence tomography (OCT) system for studying the stratum corneum in non-palmar human skin. An axial resolution of 1 µm in tissue and at least −75 dB sensitivity have been achieved. High-quality B-scans, containing...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Revin, D.G, Byers, R.A, Duan, M.Q, Li, W, Matcher, S.J
Format: Artikel
Sprache:eng
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We demonstrate a free-space, trolley-mountable Fourier domain visible-light optical coherence tomography (OCT) system for studying the stratum corneum in non-palmar human skin. An axial resolution of 1 µm in tissue and at least −75 dB sensitivity have been achieved. High-quality B-scans, containing 1600 A-scans, are acquired at a rate of 39 Hz. Images from the dorsal hand, ventral wrist and ventral forearm areas are obtained, with a clearly resolved stratum corneum layer (typically 5–15 µm thick) presenting as a hypoechogenic dark layer below the bright entrance signal, similar to that found in palmar skin with traditional OCT systems. We find that the appearance of the stratum corneum layer strongly depends on its water content, becoming brighter after occlusive hydration.
DOI:10.1364/boe.494356