OPTIMIZATION OF A MIDINFRARED HIGH-RESOLUTION DIFFERENCE-FREQUENCY LASER SPECTROMETER

Improved performance of a continuous-wave (cw) laser spectrometer for Doppler-limited infrared spectroscopy of molecules based on difference-frequency generation (DFG) in AgGaS2 has been achieved. The spectrometer was configured to generate continuous scans of up to 1 cm-1 from 1550 to 2100 cm-1. An...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of the Optical Society of America. B, Optical physics Optical physics, 1992-11, Vol.9 (11), p.1962-1967
Hauptverfasser: HIELSCHER, AH, MILLER, CE, BAYARD, DC, SIMON, U, SMOLKA, KP, CURL, RF, TITTEL, FK
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Improved performance of a continuous-wave (cw) laser spectrometer for Doppler-limited infrared spectroscopy of molecules based on difference-frequency generation (DFG) in AgGaS2 has been achieved. The spectrometer was configured to generate continuous scans of up to 1 cm-1 from 1550 to 2100 cm-1. An absolute precision of approximately 6 x 10(-3) cm-1 with a resolution of better than 1.0 MHz (3.3 x 10(-5) cm-1) was achieved. Infrared powers of approximately 20 muW were obtained by employing 90-degrees Type I phase matching in a 45-mm-long AgGaS2 crystal. The high-resolution characteristics of the DFG spectrometer were evaluated by using H2O and N2O spectra. Preliminary infrared kinetic spectroscopy results involving the detection of transient CO radicals from 193-nm acetone photodissociation are also reported.
ISSN:0740-3224
1520-8540
DOI:10.1364/JOSAB.9.001962