Initial clinical experience of N13-ammonia myocardial perfusion PET/CT using a compact superconducting production system

Although N13-ammonia has favorable properties among FDA approved radiotracers, complexity of implementation has limited its use. We describe the initial patient experience of N13-ammonia PET imaging using a compact N13-ammonia production system. N13 was produced using the ION-12SC, a 12MeV, 10uA sup...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of nuclear cardiology 2021-02, Vol.28 (1), p.295-299
Hauptverfasser: Pieper, Justin, Patel, Vaiibhav N., Escolero, Sylvia, Nelson, Jacob R., Poitrasson-Rivière, Alexis, Shreves, Christopher K., Freiburger, Nick, Hubers, David, Rothley, Jill, Corbett, James R., Oliverio, Joseph, Ficaro, Edward P., Weinberg, Richard L., Murthy, Venkatesh L.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Although N13-ammonia has favorable properties among FDA approved radiotracers, complexity of implementation has limited its use. We describe the initial patient experience of N13-ammonia PET imaging using a compact N13-ammonia production system. N13 was produced using the ION-12SC, a 12MeV, 10uA superconducting minimally shielded cyclotron, and reduced to N13-ammonia in an automated multi-use purification unit. Patients were power injected with 9.3 ± 1.1 mCi (344.1 ± 40.7 MBq) of N13-ammonia for rest imaging, and 18.8 ± 0.9 mCi (695.6 ± 33.3 MBq) of N13-ammonia was injected at peak hyperemia for stress testing. Images were interpreted for relative perfusion, left ventricular volumes/function, blood flow quantification, and scored for image quality. In total 97 patients underwent 98 N13-ammonia PET scans (32 rest only/65 rest-stress/1 stress only). Image quality was 91.8% good or excellent. None were poor/non-diagnostic. Study durations were acceptable. Tracer related radiation dosimetry to patients was 0.7 ± 0.1 mSv (rest only), and 2.1 ± 0.1 mSv (rest-stress). Clinical N13-ammonia production by the Ionetix ION-12SC delivers high quality myocardial PET perfusion images in a rapid protocol.
ISSN:1071-3581
1532-6551
DOI:10.1007/s12350-019-01886-7