Incidence, Characteristics, Predictors, and Outcomes of Surgical Explantation After Transcatheter Aortic Valve Replacement
Currently, there is a paucity of information on surgical explantation after transcatheter aortic valve replacement (TAVR). The purpose of this study was to examine the incidence, patient characteristics, predictors, and outcomes of surgical explantation after TAVR using a population-based, nationall...
Gespeichert in:
Veröffentlicht in: | Journal of the American College of Cardiology 2020-10, Vol.76 (16), p.1848-1859 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Currently, there is a paucity of information on surgical explantation after transcatheter aortic valve replacement (TAVR).
The purpose of this study was to examine the incidence, patient characteristics, predictors, and outcomes of surgical explantation after TAVR using a population-based, nationally representative database.
We analyzed the Medicare Provider profile to include all U.S. patients undergoing TAVR from 2012 to 2017. Time to surgical explant was calculated from the index TAVR discharge to surgical explantation. Post-operative survival was assessed using time-dependent Cox proportional hazard regression analysis and landmark analysis.
The incidence of surgical explantation was 0.2% (227 of 132,633 patients), and was 0.28% and 0.14% in the early and newer TAVR era, respectively. The median time to surgical explant was 212 days, whereas 8.8% and 70.9% underwent surgical explantation within 30 days and 1 year, respectively. The primary indication for reintervention was bioprosthetic failure (79.3%). Compared with the no-explant cohort, the explant cohort was significantly younger (mean age 73.7 years vs. 81.7 years), with a lower prevalence of heart failure (55.9% vs. 65.8%) but more likely a lower-risk profile cohort (15% vs. 2.4%; all p 0.05). The time-dependent Cox regression analysis demonstrated a higher mortality in those with surgical explantation (hazard ratio: 4.03 vs. no-explant group; 95% confidence interval: 1.81 to 8.98). Indication, time-to-surgical-explant, and year of surgical explantation were not associated with worse post-explantation survival (all p > 0.05).
The present study provides updated evidence on the incidence, timing, and outcomes of surgical explantation of a TAVR prosthesis. Although the overall incidence was low, short-term mortality was high. These findings stress the importance of future mechanistic studies on TAVR explantation and may have implications on lifetime management of aortic stenosis, particularly in younger patients.
[Display omitted] |
---|---|
ISSN: | 0735-1097 1558-3597 |
DOI: | 10.1016/j.jacc.2020.08.048 |