Predicting the Farnsworth-Munsell D15 and Holmes-Wright-A lantern outcomes with computer-based color vision tests
This study determined the AC1 agreement values between computer-based color vision tests and the Farnworth- Munsell D-15 (F-D15) and the Holmes-Wright Type A lantern (HWA). The computer-based tests were the United States Air Force Cone Contrast Test (OCCT), Cambridge Color Test, Innova Rabin Cone Co...
Gespeichert in:
Veröffentlicht in: | Journal of the Optical Society of America. A, Optics, image science, and vision Optics, image science, and vision, 2020-04, Vol.37 (4), p.A1-A10 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This study determined the AC1 agreement values between computer-based color vision tests and the Farnworth- Munsell D-15 (F-D15) and the Holmes-Wright Type A lantern (HWA). The computer-based tests were the United States Air Force Cone Contrast Test (OCCT), Cambridge Color Test, Innova Rabin Cone Contrast, Konan- Waggoner D15 (KWC-D15), and Color Assessment and Diagnosis (CAD). Sixty-eight color-vision-defective persons participated. The KWC-D15 had the highest AC1 with the F-D15 (AC1 = 0.88). Both the CAD and OCCT had the highest values with the HWA (AC1 > 0.96). The KWC-D15 would be the best substitute for the F-D15. Either the CAD or OCCT would be appropriate substitutes for the HWA. (C) 2020 Optical Society of America |
---|---|
ISSN: | 1084-7529 1520-8532 |
DOI: | 10.1364/JOSAA.381305 |