Study of a multi-wire proportional chamber with a cathode strip and delay-line readout

The design principle for a multi-wire proportional chamber with a cathode strip and delay-line readout is described. A prototype chamber of a size of 10 cm×10 cm was made together with the readout electronics circuit. A very clean signal with very low background noise was obtained by applying a tran...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Chinese physics C 2009-05, Vol.33 (5), p.364-368
1. Verfasser: 韩丽英 李奇特 Q. Faisal 葛愉成 刘洪涛 叶沿林
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The design principle for a multi-wire proportional chamber with a cathode strip and delay-line readout is described. A prototype chamber of a size of 10 cm×10 cm was made together with the readout electronics circuit. A very clean signal with very low background noise was obtained by applying a transformer between the delay-line and the pre-amplifier in order to match the resistance. Along the anode wire direction a position resolution of less than 0.5 mm was achieved with a ^55Fe-5.9 keV X ray source. The simple structure, large effective area and high position resolution allow the application of a gas chamber of this kind to many purposes.
ISSN:1674-1137
0254-3052
2058-6132
DOI:10.1088/1674-1137/33/5/009