Nonlinear dynamic response and stability analysis of the stapes reconstruction in human middle ear

Stapes fracture causes hearing loss and instability in the middle ear hearing system (MEHS). The material used in the stapes reconstruction restores stapes, but the effects of the nonlinear material parameters on the stability of the MEHS are still unknown. To address this challenge, the nonlinear d...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Applied mathematics and mechanics 2023-10, Vol.44 (10), p.1739-1760
Hauptverfasser: Wang, Lele, Wang, Liang, Zhu, Yueting, Liu, Zhanli, Sun, Yongtao, Wang, Jie, Han, Hongge, Xiang, Shuyi, Shi, Huibin, Ding, Qian
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Stapes fracture causes hearing loss and instability in the middle ear hearing system (MEHS). The material used in the stapes reconstruction restores stapes, but the effects of the nonlinear material parameters on the stability of the MEHS are still unknown. To address this challenge, the nonlinear dynamic response and stability of the stapes reconstruction are investigated using a multi-degree-of-freedom mechanical model. The material parameters of the implant are tentatively determined by analyzing the natural frequencies of the undamped system. The dynamical properties of the MEHS are characterized under different external excitations. The approximate solution of the MEHS near the resonant frequency is derived through the multiple-time-scale method (MTSM). The results show that the nonlinear stiffness of the material has little influence on the MEHS in the healthy state, but it causes resonant phenomena between the ossicle and the implant in the pathological state.
ISSN:0253-4827
1573-2754
DOI:10.1007/s10483-023-3037-9