An HIV1/2 point of care test on sputum for screening TB/HIV co-infection in central India – Will it work?

To determine whether the OraQuick® HIV-1/2 Assay (OraSure Technologies, Inc., Bethlehem, PA, USA) in sputum is a valid tool for HIV surveillance among TB patients. A cross sectional study was carried out on sputa of patients diagnosed with tuberculosis. Sputa were tested for antibodies to HIV using...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Asian Pacific journal of tropical medicine 2013-03, Vol.6 (3), p.216-219
Hauptverfasser: Desikan, Prabha, De, Sajal, Pai, Nitika Pant, Mishra, Pradyumna K, Kumar, Kaushal, Panwalkar, Nikita, Verma, Mayanka, Hasan, Zia Ul, Maudar, Kewal K
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:To determine whether the OraQuick® HIV-1/2 Assay (OraSure Technologies, Inc., Bethlehem, PA, USA) in sputum is a valid tool for HIV surveillance among TB patients. A cross sectional study was carried out on sputa of patients diagnosed with tuberculosis. Sputa were tested for antibodies to HIV using OraQuick® HIV-1/2 Assay (OraSure Technologies, Inc., Bethlehem, PA, USA). The results were compared with results of serum ELISA. Compared to serum ELISA, the OraQuick® HIV-1/2 Assay in sputum specimens reported 90% sensitivity (9/10) and 100% specificity (307/307), with a positive predictive value of 100% (95%CI: 66.37%–100.00%) and a negative predictive value of 99.68% (95%CI: 98.20%–99.99%). This testing method may provide a useful strategy for conducting HIV surveillance in possible co-infected TB patients at peripheral centres. Since there is no investment on infrastructure, it may be possible for paramedical health professionals to carry out the test, particularly in areas with low HIV endemicity.
ISSN:1995-7645
2352-4146
DOI:10.1016/S1995-7645(13)60026-7