Two-phase flow component fraction measurement using gamma-ray attenuation technique

Multiphase flow meters as the potential alternatives to separation and metering techniques have been in rapid development since 1980s. Before its field operation, the instrument should be calibrated in a standard test-facility. In spite of the known medium and large scale facilities all over the wor...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Nuclear science and techniques 2017-06, Vol.28 (6), p.196-202, Article 88
Hauptverfasser: Sharifzadeh, Mohsen, Khalafi, Hosein, Afarideh, Hosein, Noori, Ehsanallah
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Multiphase flow meters as the potential alternatives to separation and metering techniques have been in rapid development since 1980s. Before its field operation, the instrument should be calibrated in a standard test-facility. In spite of the known medium and large scale facilities all over the world, we developed a laboratory scale instrument for component fraction measurements. It has a two-phase flow homogenizer loop with the clamp-on potential of the meters to provide a regime independent measurement. It is capable of delivering a complete homogenization by c-ray densitometer. With an error of ±5% in component fraction measurements, this instrument is appropriate for testing and calibrating other meters.
ISSN:1001-8042
2210-3147
DOI:10.1007/s41365-017-0237-4