Determination of Xenon in Air by a Pulse-discharge Helium Ionization Detector

A pulse-discharge helium ionization detector(Valco, PD-D3-I) was used to measure xenon concentration in air. The dependences of the detector relative response on various gas chromatograph parameters were investigated. Based on the well prepared gas connections for the detector system and optimized g...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:高等学校化学研究(英文版) 2012-09, Vol.28 (5), p.814-817
1. Verfasser: CHEN Zhan-ying CHANG Yin-zhong LIU Shu-jiang ZHANG Hai-tao WANG Shi-lian LI Qi
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A pulse-discharge helium ionization detector(Valco, PD-D3-I) was used to measure xenon concentration in air. The dependences of the detector relative response on various gas chromatograph parameters were investigated. Based on the well prepared gas connections for the detector system and optimized gas chromatography(GC) working conditions, the atmospheric xenon concentration could be measured by the cheap GC method with a detection level of 0.7×10-9(parts by volume). Moreover, the xenon concentration in the ground level air around our laboratory was measured with the result of 0.085×10-6(parts by volume) and RSD of 0.91%.
ISSN:1005-9040
2210-3171