A modified extraction protocol enables detection and quantification of celiac disease-related gluten proteins from wheat

The detection, analysis, and quantification of individual celiac disease (CD) immune responsive gluten proteins in wheat and related cereals (barley, rye) require an adequate and reliable extraction protocol. Because different types of gluten proteins behave differently in terms of solubility, curre...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of chromatography. B, Analytical technologies in the biomedical and life sciences Analytical technologies in the biomedical and life sciences, 2009-04, Vol.877 (10), p.975-982
Hauptverfasser: van den Broeck, Hetty C., America, Antoine H.P., Smulders, Marinus J.M., Bosch, Dirk, Hamer, Rob J., Gilissen, Ludovicus J.W.J., van der Meer, Ingrid M.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The detection, analysis, and quantification of individual celiac disease (CD) immune responsive gluten proteins in wheat and related cereals (barley, rye) require an adequate and reliable extraction protocol. Because different types of gluten proteins behave differently in terms of solubility, currently different extraction protocols exist. The performance of various documented gluten extraction protocols is evaluated for specificity and completeness by gel electrophoresis (SDS-PAGE), immunoblotting and RIDASCREEN ® Gliadin competitive ELISA. Based on these results, an optimized, two-step extraction protocol has been developed.
ISSN:1570-0232
1873-376X