Single biosensor immunoassay for the detection of five aminoglycosides in reconstituted skimmed milk
The application of an optical biosensor (Biacore 3000), with four flow channels (Fcs), in combination with a mixture of four specific antibodies resulted in a competitive inhibition biosensor immunoassay (BIA) for the simultaneous detection of the five relevant aminoglycosides in reconstituted skimm...
Gespeichert in:
Veröffentlicht in: | Analytica chimica acta 2003-07, Vol.488 (1), p.53-60 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The application of an optical biosensor (Biacore 3000), with four flow channels (Fcs), in combination with a mixture of four specific antibodies resulted in a competitive inhibition biosensor immunoassay (BIA) for the simultaneous detection of the five relevant aminoglycosides in reconstituted skimmed milk. Four aminoglycosides (gentamicin, neomycine, kanamycin and a streptomycin derivative) were immobilised onto the sensor surface of a biosensor chip (CM5) in the four Fcs of the biosensor system by amine coupling. In the Biacore, milk (reconstituted from skimmed milk powder) was 10 times diluted with a mixture of the four specific antibodies and injected through the four serially connected Fcs (1
min at a flow rate of 20
μl
min
−1). The responses measured just prior to the injection (20
μl at a flow rate of 20
μl
min
−1) of the regeneration solution (0.2
M NaOH + 20% acetonitril) were indicative for the presence or absence of the aminoglycosides in reconstituted milk. The limits of detection were between 15 and 60
ng
ml
−1, which was far below the maximum residue limits (MRLs) (varying from 100 to 500
ng
ml
−1) and the total run time between samples was 7
min. |
---|---|
ISSN: | 0003-2670 1873-4324 |
DOI: | 10.1016/S0003-2670(03)00628-7 |