A new method for epidural catheter fixation

We investigated the effect of a new fixator made from a medication port of a intravenous fluid container on the migration of epidural catheter. Fifty patients were recruited to receive epidural analgesia and allocated to one of two groups. In the new fixator group (n=25) epidural catheter was advanc...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Aǧrı 2007-04, Vol.19 (2), p.33-37
Hauptverfasser: Gülcü, Nebahat, Karaaslan, Kazim, Koçoğlu, Hasan, Gümüş, Esra
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We investigated the effect of a new fixator made from a medication port of a intravenous fluid container on the migration of epidural catheter. Fifty patients were recruited to receive epidural analgesia and allocated to one of two groups. In the new fixator group (n=25) epidural catheter was advanced through a fixator then fixed with transparent adhesive dressing, in the standard dressing group (n=25), the catheter was fixed only with transparent adhesive dressing. Outward migration of the catheter over 2 cm, and inward migration over 1 cm was described as the failure of the fixation. At the end of the follow up time, outward migration or inward migration distance, and the reason for removing the catheter were recorded. In the standard dressing group, outward migration was detected in 7 patients compared to 2 patients in the new fixator group (p>0.05). Inward migration was determined in only 5 cases in the standard dressing group (p
ISSN:1300-0012