A fully automated one-pot synthesis of [carbonyl- 11C]WAY-100635 for clinical PET applications
A fully automated synthesis of the important 5HT 1A receptor radioligand, [carbonyl- 11C]WAY-100635 ( I), was developed based on the optimized one-pot “wet” synthesis procedure. A modern automated apparatus was constructed from commercially available components and operated via LabView software. In...
Gespeichert in:
Veröffentlicht in: | Applied radiation and isotopes 2009, Vol.67 (1), p.73-78 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A fully automated synthesis of the important 5HT
1A receptor radioligand, [carbonyl-
11C]WAY-100635 (
I), was developed based on the optimized one-pot “wet” synthesis procedure. A modern automated apparatus was constructed from commercially available components and operated via LabView software. In average, (906±525)
MBq (
n=94) of (
I) was obtained from 40
min bombardment at 50
μA beam current within 50
min synthesis time. The specific radioactivity (SA) at the time of injection was (50.5±29.3)
GBq/μmol (
n=94). |
---|---|
ISSN: | 0969-8043 1872-9800 1872-9800 |
DOI: | 10.1016/j.apradiso.2008.07.008 |