Development and validation of non‐guided bladder‐neck and neurovascular‐bundle dissection modules of the RobotiX‐Mentor® full‐procedure robotic‐assisted radical prostatectomy virtual reality simulation
Background Full‐procedure virtual reality (VR) simulator training in robotic‐assisted radical prostatectomy (RARP) is a new tool in surgical education. Methods Description of the development of a VR RARP simulation model, (RobotiX‐Mentor®) including non‐guided bladder neck (ngBND) and neurovascular...
Gespeichert in:
Veröffentlicht in: | The international journal of medical robotics + computer assisted surgery 2021-04, Vol.17 (2), p.e2195-n/a |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background
Full‐procedure virtual reality (VR) simulator training in robotic‐assisted radical prostatectomy (RARP) is a new tool in surgical education.
Methods
Description of the development of a VR RARP simulation model, (RobotiX‐Mentor®) including non‐guided bladder neck (ngBND) and neurovascular bundle dissection (ngNVBD) modules, and assessment of face, content, and construct validation of the ngBND and ngNVBD modules by robotic surgeons with different experience levels.
Results
Simulator and ngBND/ngNVBD modules were rated highly by all surgeons for realism and usability as training tool. In the ngBND‐task construct, validation was not achieved in task‐specific performance metrics. In the ngNVBD, task‐specific performance of the expert/intermediately experienced surgeons was significantly better than that of novices.
Conclusions
We proved face and content validity of simulator and both modules, and construct validity for generic metrics of the ngBND module and for generic and task‐specific metrics of the ngNVBD module. |
---|---|
ISSN: | 1478-5951 1478-596X |
DOI: | 10.1002/rcs.2195 |