Mechanical extraction of cardiac implantable electronic devices leads with long dwell time: Efficacy and safety of the step up approach
The aim of this study was to evaluate the efficacy and safety of the stepwise mechanical transvenous lead extraction approach in a patient population with chronically implanted transvenous leads with a long dwell time. From January 2014 till December 2018, all lead extractions with lead dwell time ≥...
Gespeichert in:
Veröffentlicht in: | Pacing and clinical electrophysiology 2021-01, Vol.44 (1), p.120-128 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The aim of this study was to evaluate the efficacy and safety of the stepwise mechanical transvenous lead extraction approach in a patient population with chronically implanted transvenous leads with a long dwell time.
From January 2014 till December 2018, all lead extractions with lead dwell time ≥5 years performed at our tertiary centre were retrospectively analysed. A total of 173 leads, from 78 patients (median age 68 years; 81% male) with a median dwell time of 9 years (interquartile range [IQR] 5) were extracted, with three or more leads in 42% of the patients. Right atrial leads: 41%; right ventricular pacing leads: 16%; implantable cardioverter‐defibrillator (ICD) leads: 31% (72% dual coil); coronary sinus leads: 12%. The majority (75%) of the leads had an active fixation. Most frequent indication for extraction was pocket infection/erosion (76%). Overall clinical success was 97%, and complete procedural success was 93%. Venous patency, assessed with venous angiography, was well preserved in 93% of the cases. The overall procedural complication rate was 3.8% (2.6% major and 1.3% minor).
Despite the complexity of the population and a very long dwell time (median 9 years), a clinical success rate of 97% was achieved with the stepwise mechanical approach. Analysis of impeding progression of pectoral extraction suggests that dense fibrosis and sharp lead curvature in the transvenous trajectory pose a challenge. Complication rate was low, and acute venous patency was generally well preserved. |
---|---|
ISSN: | 0147-8389 1540-8159 |
DOI: | 10.1111/pace.14094 |