End-expiratory lung volume remains stable during N 2 MBW in healthy sleeping infants
We have previously shown that functional residual capacity (FRC) and lung clearance index were significantly greater in sleeping healthy infants when measured by N (nitrogen) washout using 100% O (oxygen) versus 4% SF (sulfur hexafluoride) washout using air. Following 100% O exposure, tidal volumes...
Gespeichert in:
Veröffentlicht in: | Physiological reports 2020-08, Vol.8 (16), p.e14477 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We have previously shown that functional residual capacity (FRC) and lung clearance index were significantly greater in sleeping healthy infants when measured by N
(nitrogen) washout using 100% O
(oxygen) versus 4% SF
(sulfur hexafluoride) washout using air. Following 100% O
exposure, tidal volumes decreased by over 30%, while end-expiratory lung volume (EELV, i.e., FRC) rose markedly based on ultrasonic flow meter assessments. In the present study to investigate the mechanism behind the observed changes, N
MBW was performed in 10 separate healthy full-term spontaneously sleeping infants, mean (range) 26 (18-31) weeks, with simultaneous EELV monitoring (respiratory inductance plethysmography, RIP) and oxygen uptake (V´O
) assessment during prephase air breathing, during N
washout by exposure to 100% O
, and subsequently during air breathing. While flow meter signals suggested a rise in ELLV by mean (SD) 26 (9) ml over the washout period, RIP signals demonstrated no EELV change. V'O
/FRC ratio during air breathing was mean (SD) 0.43 (0.08)/min, approximately seven times higher than that calculated from adult data. We propose that our previously reported flow meter-based overestimation of EELV was in fact a physiological artifact caused by rapid and marked movement of O
across the alveolar capillary membrane into the blood and tissue during 100% O
exposure, without concomitant transfer of N
to the same degree in the opposite direction. This may be driven by the high observed O
consumption and resulting cardiac output encountered in infancy. Furthermore, the low resting lung volume in infancy may make this error in lung volume determination by N
washout relatively large. |
---|---|
ISSN: | 2051-817X 2051-817X |
DOI: | 10.14814/phy2.14477 |