Accuracy of spectral Doppler flow and tissue velocity measurements in ultrasound systems

Blood and tissue velocity are measured and analysed in cardiac, vascular and other applications of diagnostic ultrasound (US). An error in system calibration is a potential risk for misinterpretation of the measurements. To determine the accuracy in velocity calibration, we tested three common comme...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Ultrasound in medicine & biology 2004, Vol.30 (1), p.127-132
Hauptverfasser: Walker, Andrew, Olsson, Eva, Wranne, Bengt, Ringqvist, Ivar, Ask, Per
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Blood and tissue velocity are measured and analysed in cardiac, vascular and other applications of diagnostic ultrasound (US). An error in system calibration is a potential risk for misinterpretation of the measurements. To determine the accuracy in velocity calibration, we tested three common commercial US systems using a Doppler string phantom. We tested pulsed and continuous-wave Doppler modes for velocities relevant to both cardiac blood flow and tissue-velocity estimation. The US systems were tested with settings and transducers commonly used in cardiac applications. One system consistently overestimated velocity by about 5%, whereas the other two systems were quite accurate in velocity estimation. These findings emphasize the importance of continuous quality control of US equipment. (E-mail: andrew.walker@ltvastmanland.se)
ISSN:0301-5629
1879-291X
1879-291X
DOI:10.1016/j.ultrasmedbio.2003.08.020