Accuracy of spectral Doppler flow and tissue velocity measurements in ultrasound systems
Blood and tissue velocity are measured and analysed in cardiac, vascular and other applications of diagnostic ultrasound (US). An error in system calibration is a potential risk for misinterpretation of the measurements. To determine the accuracy in velocity calibration, we tested three common comme...
Gespeichert in:
Veröffentlicht in: | Ultrasound in medicine & biology 2004, Vol.30 (1), p.127-132 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Blood and tissue velocity are measured and analysed in cardiac, vascular and other applications of diagnostic ultrasound (US). An error in system calibration is a potential risk for misinterpretation of the measurements. To determine the accuracy in velocity calibration, we tested three common commercial US systems using a Doppler string phantom. We tested pulsed and continuous-wave Doppler modes for velocities relevant to both cardiac blood flow and tissue-velocity estimation. The US systems were tested with settings and transducers commonly used in cardiac applications. One system consistently overestimated velocity by about 5%, whereas the other two systems were quite accurate in velocity estimation. These findings emphasize the importance of continuous quality control of US equipment. (E-mail:
andrew.walker@ltvastmanland.se) |
---|---|
ISSN: | 0301-5629 1879-291X 1879-291X |
DOI: | 10.1016/j.ultrasmedbio.2003.08.020 |