High‐precision determination of lithium and magnesium isotopes utilising single column separation and multi‐collector inductively coupled plasma mass spectrometry
Rationale Li and Mg isotopes are increasingly used as a combined tool within the geosciences. However, established methods require separate sample purification protocols utilising several column separation procedures. This study presents a single‐step cation‐exchange method for quantitative separati...
Gespeichert in:
Veröffentlicht in: | Rapid communications in mass spectrometry 2018-01, Vol.32 (2), p.93-104 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Rationale
Li and Mg isotopes are increasingly used as a combined tool within the geosciences. However, established methods require separate sample purification protocols utilising several column separation procedures. This study presents a single‐step cation‐exchange method for quantitative separation of trace levels of Li and Mg from multiple sample matrices.
Methods
The column method utilises the macro‐porous AGMP‐50 resin and a high‐aspect ratio column, allowing quantitative separation of Li and Mg from natural waters, sediments, rocks and carbonate matrices following the same elution protocol. High‐precision isotope determination was conducted by multi‐collector inductively coupled plasma mass spectrometry (MC‐ICPMS) on the Thermo Scientific™ NEPTUNE Plus™ fitted with 1013 Ω amplifiers which allow accurate and precise measurements at ion beams ≤0.51 V.
Results
Sub‐nanogram Li samples (0.3–0.5 ng) were regularly separated (yielding Mg masses of 1–70 μg) using the presented column method. The total sample consumption during isotopic analysis is |
---|---|
ISSN: | 0951-4198 1097-0231 1097-0231 |
DOI: | 10.1002/rcm.8020 |