Round robin study to evaluate the reconstructed human epidermis (RhE) model as an in vitro skin irritation test for detection of irritant activity in medical device extracts
Assessment of skin irritation is an essential component of the safety evaluation of medical devices. OECD Test Guideline 439 describes the use of reconstructed human epidermis (RhE) as an in vitro test system for classification of skin irritation by neat chemicals. An international round robin study...
Gespeichert in:
Veröffentlicht in: | Toxicology in vitro 2018-08, Vol.50, p.439-449 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Assessment of skin irritation is an essential component of the safety evaluation of medical devices. OECD Test Guideline 439 describes the use of reconstructed human epidermis (RhE) as an in vitro test system for classification of skin irritation by neat chemicals. An international round robin study was conducted to evaluate the RhE method for determination of skin irritant potential of medical device extracts. Four irritant polymers and three non-irritant controls were obtained or developed that had demonstrated their suitability to act as positive or negative test samples. The RhE tissues (EpiDerm™ and SkinEthic™ RHE) were dosed with 100 μL aliquots of either saline or sesame oil extract. Incubation times were 18 h (EpiDerm™) and 24 h (SkinEthic™ RHE). Cell viability reduction > 50% was indicative of skin irritation. Both the EpiDerm™ and SkinEthic™ RHE tissues were able to correctly identify virtually all of the irritant polymer samples either in the saline, sesame oil or both solvent extracts. Our results indicate that RhE tissue models can detect the presence of strong skin irritants at low levels in dilute medical device polymer extracts. Therefore, these models may be suitable replacements for the rabbit skin irritation test to support the biological evaluation of medical devices.
•Alternative in vitro testing of medical device extracts for irritation is possible.•Reconstructed human epidermis (RhE) model detects presence of irritant in extracts.•High predictivity RhE model for presence of strong irritants in material extracts.•Predictivity in saline or sesame oil extract depends on hydrophilicity of irritant.•Four positive polymer test materials containing a known irritant were evaluated. |
---|---|