Reducing the flow measurement error caused by pulsations in flows

Different types of errors are generated by pulsations in flows. Among these errors is the sampling error due to a unadapted time-averaging of the flow rate. An improved model for pulsations in flows including harmonics is derived. The localization of the harmonics is performed by a detector. The per...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Flow measurement and instrumentation 2004-10, Vol.15 (5), p.311-315
Hauptverfasser: Berrebi, J., van Deventer, J., Delsing, J.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Different types of errors are generated by pulsations in flows. Among these errors is the sampling error due to a unadapted time-averaging of the flow rate. An improved model for pulsations in flows including harmonics is derived. The localization of the harmonics is performed by a detector. The period of the pulsations is estimated. It is then possible to reduce the sampling error by performing a correct averaging. The reduction of the sampling error is confirmed by simulations.
ISSN:0955-5986
1873-6998
1873-6998
DOI:10.1016/j.flowmeasinst.2004.05.001