TTF-Modified Biosensors for Hydrogen Peroxide

Amperometric biosensors for hydrogen peroxide were produced, based on TTF-modified graphite disc electrodes, incorporating immobilised peroxidase. At the optimum operating potential (−200 mV vs. calomel electrode), a linear response to hydrogen peroxide was obtained in the range 0-2 mM (Km value = 1...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytical letters 1994-01, Vol.27 (8), p.1443-1452
Hauptverfasser: Bifulco, Laura, Cammaroto, Cosimo, Newman, Jeffrey D., Turner, Anthony P.F.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Amperometric biosensors for hydrogen peroxide were produced, based on TTF-modified graphite disc electrodes, incorporating immobilised peroxidase. At the optimum operating potential (−200 mV vs. calomel electrode), a linear response to hydrogen peroxide was obtained in the range 0-2 mM (Km value = 1.98 mM). Low sensitivity to pH fluctuations and good stability at room temperature were exhibited.
ISSN:0003-2719
1532-236X
1532-236X
DOI:10.1080/00032719408006381