Central and peripheral refraction measured by a novel double-pass instrument
A novel double-pass instrument and its data analysis method for the measurement of central and peripheral refraction is presented and validated in a group of healthy subjects. The instrument acquires in-vivo, non-cycloplegic, double-pass, through-focus images of the eye's central and peripheral...
Gespeichert in:
Veröffentlicht in: | Biomedical optics express 2023-06, Vol.14 (6), p.2608-2617 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A novel double-pass instrument and its data analysis method for the measurement of central and peripheral refraction is presented and validated in a group of healthy subjects. The instrument acquires in-vivo, non-cycloplegic, double-pass, through-focus images of the eye's central and peripheral point-spread function (PSF) using an infrared laser source, a tunable lens and a CMOS camera. The through-focus images were analyzed to determine defocus and astigmatism at 0° and 30° visual field. These values were compared to those obtained with a lab-based Hartmann-Shack wavefront sensor. The two instruments provided data showing good correlation at both eccentricities, particularly in the estimation of defocus. |
---|---|
ISSN: | 2156-7085 2156-7085 |
DOI: | 10.1364/BOE.489881 |