Application of a sulphide-selective electrode in the absence of a pH-buffer
Calibration of a sulphide electrode in the pH-range 9–12 has been studied as an e.m.f. vs. (pH — p[HS −]) function by measuring e.m.f. and pH in parallel. Calibration can also be done in this pH range by using a differential amplifier with a three-electrode measuring cell (glass, sulphide-selective...
Gespeichert in:
Veröffentlicht in: | Talanta (Oxford) 1988-10, Vol.35 (10), p.783-788 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Calibration of a sulphide electrode in the pH-range 9–12 has been studied as an e.m.f.
vs. (pH — p[HS
−]) function by measuring e.m.f. and pH in parallel. Calibration can also be done in this pH range by using a differential amplifier with a three-electrode measuring cell (glass, sulphide-selective and reference electrodes). The effects of an antioxidant (ascorbic acid) and a complexing agent (DCTA) on the calibration of the glass-sulphide electrode cell at pH < 5 were studied. The applicability of this end-point indicator cell has been demonstrated for titrations of Ag
+, Pb
2+ and Bi
3+ with Na
2S. |
---|---|
ISSN: | 0039-9140 1873-3573 1873-3573 |
DOI: | 10.1016/0039-9140(88)80183-8 |