A photon-counting silicon-strip detector for digital mammography with an ultrafast 0.18-μm CMOS ASIC

We have evaluated a silicon-strip detector with a 0.18-μm CMOS application specific integrated circuits (ASIC) containing 160 channels for use in photon-counting digital mammography. Measurements were performed at the Elettra light source using monochromatic X-ray beams with different energies and i...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Accelerators, spectrometers, detectors and associated equipment, 2014-06, Vol.749, p.1-6
Hauptverfasser: Chen, Han, Cederström, Björn, Xu, Cheng, Persson, Mats, Karlsson, Staffan, Danielsson, Mats
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We have evaluated a silicon-strip detector with a 0.18-μm CMOS application specific integrated circuits (ASIC) containing 160 channels for use in photon-counting digital mammography. Measurements were performed at the Elettra light source using monochromatic X-ray beams with different energies and intensities. Energy resolution, ΔE/Ein, was measured to vary between 0.10 and 0.23 in the energy range of 15–40keV. Pulse pileup has shown little effect on energy resolution.
ISSN:0168-9002
1872-9576
1872-9576
DOI:10.1016/j.nima.2014.02.033