An integrated QCM-based narcotics sensing microsystem

We present the design, fabrication and successful testing of a 14x14x4 mm3 integrated electronic narcotics sensing system which consists of only four parts. The microsystem absorbs airborne narcotics molecules and performs a liquid assay using an integrated quartz crystal microbalance (QCM). A verti...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Lab on a chip 2008, Vol.8 (10), p.1648-1657
Hauptverfasser: Frisk, Thomas, Sandström, Niklas, Eng, Lars, van der Wijngaart, Wouter, Månsson, Per, Stemme, Göran
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We present the design, fabrication and successful testing of a 14x14x4 mm3 integrated electronic narcotics sensing system which consists of only four parts. The microsystem absorbs airborne narcotics molecules and performs a liquid assay using an integrated quartz crystal microbalance (QCM). A vertically conductive double-sided adhesive foil (VCAF) was used and studied as a novel material for LOC and MEMS applications and provides easy assembly, electrical contacting and liquid containment. The system was tested for measuring cocaine and ecstasy, with successful detection of amounts as small as 100 ng and 200 ng, respectively. These levels are of interest in security activities in customs, prisons and by the police.
ISSN:1473-0197
1473-0189
1473-0189
DOI:10.1039/b800487k