Optimizing the dynamic range extension of a radiochromic film dosimetry system
The authors present a radiochromic film dosimetry protocol for a multicolor channel radiochromic film dosimetry system consisting of the external beam therapy (EBT) model GAFCHROMIC™ film and the Epson Expression 1680 flat-bed document scanner. Instead of extracting only the red color channel, the a...
Gespeichert in:
Veröffentlicht in: | Medical physics (Lancaster) 2009-02, Vol.36 (2), p.429-437 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The authors present a radiochromic film dosimetry protocol for a multicolor channel radiochromic film dosimetry system consisting of the external beam therapy (EBT) model GAFCHROMIC™ film and the Epson Expression 1680 flat-bed document scanner. Instead of extracting only the red color channel, the authors are using all three color channels in the absorption spectrum of the EBT film to extend the dynamic dose range of the radiochromic film dosimetry system. By optimizing the dose range for each color channel, they obtained a system that has both precision and accuracy below 1.5%, and the optimized ranges are
0
–
4
Gy
for the red channel,
4
–
50
Gy
for the green channel, and above
50
Gy
for the blue channel. |
---|---|
ISSN: | 0094-2405 2473-4209 |
DOI: | 10.1118/1.3049597 |