Properties of unflattened photon beams shaped by a multileaf collimator

Several studies have shown that removal of the flattening filter from the treatment head of a clinical accelerator increases the dose rate and changes the lateral profile in radiation therapy with photons. However, the multileaf collimator (MLC) used to shape the field was not taken into considerati...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical physics (Lancaster) 2006-05, Vol.33 (6), p.1738-1746
Hauptverfasser: Pönisch, Falk, Titt, Uwe, Vassiliev, Oleg N., Kry, Stephen F., Mohan, Radhe
Format: Artikel
Sprache:jpn
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Several studies have shown that removal of the flattening filter from the treatment head of a clinical accelerator increases the dose rate and changes the lateral profile in radiation therapy with photons. However, the multileaf collimator (MLC) used to shape the field was not taken into consideration in these studies. We therefore investigated the effect of the MLC on flattened and unflattened beams. To do this, we performed measurements on a Varian Clinac 21EX and MCNPX Monte Carlo simulations to analyze the physical properties of the photon beam. We compared lateral profiles, depth dose curves, MLC leakages, and total scatter factors for two energies (6 and 18 MV ) of MLC-shaped fields and jaw-shaped fields. Our study showed that flattening filter-free beams shaped by a MLC differ from the jaw-shaped beams. Similar differences were also observed for flattened beams. Although both collimating methods produced identical depth dose curves, the penumbra size and the MLC leakage were reduced in the softer, unflattened beam and the total scatter factors showed a smaller field size dependence.
ISSN:0094-2405
2473-4209
DOI:10.1118/1.2201149