Improved computed radiography image quality from a BaFI:Eu photostimulable phosphor plate
Recent advances in the design of photostimulable phosphor (PSP) plates for computed radiography (CR) systems have made it possible to manufacture plates made of BaFI:Eu phosphor material in the cassette form. The image quality of this plate, six BaF(Br,I):Eu plates, and one BaFBr:Eu plate were evalu...
Gespeichert in:
Veröffentlicht in: | Medical physics (Lancaster) 2002-04, Vol.29 (4), p.592-597 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Recent advances in the design of photostimulable phosphor (PSP) plates for computed radiography (CR) systems have made it possible to manufacture plates made of BaFI:Eu phosphor material in the cassette form. The image quality of this plate, six BaF(Br,I):Eu plates, and one BaFBr:Eu plate were evaluated in terms of presampling modulation transfer function (MTF), normalized Wiener spectra
(
WS
N
),
and detective quantum efficiency (DQE). Compared with the best-performing BaF(Br,I):Eu plate, the BaFI:Eu plate provided DQE that was higher, at spatial frequencies of 0.5, 1.0, and 2.0 cycles/mm, by 12% (21.8 versus 19.4), 13% (18.8 versus 16.7), and 11% (12.0 versus 10.8), respectively. Since presampling MTF values of the two plates were comparable, the BaFI:Eu plate’s higher DQE is attributable to total
WS
N
conversely being lower by 17%
[8.65×10
−5
(
mm
2
)
versus
10.38×10
−5
(
mm
2
)],
17%
[5.85×10
−5
(
mm
2
)
versus
7.03×10
−5
(
mm
2
)],
and 12%
[2.82×10
−5
(
mm
2
)
versus
3.19×10
−5
(
mm
2
)]
at the specified frequencies, respectively, primarily due to the contribution of x-ray quantum
WS
N
.
This jibes with the high x-ray absorption provided by the 27%-higher x-ray attenuation coefficients (7.54 versus 6.07, at 60 KeV) that BaFI offers over
BaF
(
Br
0.85
,
I
0.15
),
a result of the high atomic number of BaFI’s exclusively iodine halide content. The results were consistent with earlier studies of several of these same plates and indicate that BaFI:Eu is a promising avenue to lower image noise and higher overall CR system image quality. |
---|---|