Sagnac interferometer for time-resolved magneto-optical measurements

We introduce a time-resolved magneto-optical measurement technique based on a zero-area Sagnac interferometer. By replacing a continuous wave light source to a pulsed one, temporal resolution of hundreds of picoseconds is achieved. Because two lights passing through a Sagnac loop always travel the s...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of scientific instruments 2022-01, Vol.93 (1), p.013903-013903
Hauptverfasser: Heo, Hyeokjun, Kim, Taeho, Jeong, Yungi, Park, Hangyeol, Jang, Joonho
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We introduce a time-resolved magneto-optical measurement technique based on a zero-area Sagnac interferometer. By replacing a continuous wave light source to a pulsed one, temporal resolution of hundreds of picoseconds is achieved. Because two lights passing through a Sagnac loop always travel the same optical path length, the interference from the phase modulation and Kerr rotation occurs in a pulse mode. For illustration of the apparatus, we present ferromagnetic resonance of a Permalloy film caused by a magnetic field pump. The instrument still possesses the favorable properties of a Sagnac interferometer, such as rejection of all the reciprocal effects, and shows 1μrad/Hz sensitivity at a 3 µW optical power in the pulse mode.
ISSN:0034-6748
1089-7623
DOI:10.1063/5.0073908