A 4He vector zero-field optically pumped magnetometer operated in the Earth-field
Low intrinsic noise, high bandwidth, and high accuracy vector magnetometers are key components for many ground or space geophysical applications. Here, we report the design and the test of a 4He vector optically pumped magnetometer specifically dedicated to these needs. It is based on a parametric r...
Gespeichert in:
Veröffentlicht in: | Review of scientific instruments 2021-10, Vol.92 (10), p.105005-105005 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Low intrinsic noise, high bandwidth, and high accuracy vector magnetometers are key components for many ground or space geophysical applications. Here, we report the design and the test of a 4He vector optically pumped magnetometer specifically dedicated to these needs. It is based on a parametric resonance magnetometer architecture operated in the Earth magnetic field with closed-loop compensation of the three components of the magnetic field. It provides offset-free vector measurements in a ±70 μT range with a DC to 1 kHz bandwidth. We demonstrate a vector sensitivity up to 130 fT/√Hz, which is about ten times better than the best available fluxgate magnetometers currently available for the same targeted applications. |
---|---|
ISSN: | 0034-6748 1089-7623 |