Anti-contamination SMART (Spectrum Monitoring Apparatus with Roll-to-roll Transparent film) window for optical diagnostics of plasma systems
Optical emission spectroscopy is widely used in semiconductor and display manufacturing for plasma process monitoring. However, because of the contamination of the viewport, quantitative analysis is extremely difficult; therefore, qualitative analysis is used to detect species in the process. To ext...
Gespeichert in:
Veröffentlicht in: | Review of scientific instruments 2021-01, Vol.92 (1), p.013507-013507 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Optical emission spectroscopy is widely used in semiconductor and display manufacturing for plasma process monitoring. However, because of the contamination of the viewport, quantitative analysis is extremely difficult; therefore, qualitative analysis is used to detect species in the process. To extend plasma monitoring in advanced precise processes, the contamination problem of the viewport must be solved. We propose a new spectrum monitoring apparatus with a roll-to-roll transparent film window for optical diagnostics of a plasma system. By moving a transparent film in front of the viewport, contamination in the emission light path becomes negligible. However, the speed of the film should be optimized to reduce the maintenance period and to minimize measurement errors. We calculated the maximum thickness of SiO2, Si3N4, ITO, and the Ar/CHF3 plasma contaminant to suppress the electron temperature error measured by the line-intensity-ratio within 2% at 2 eV. The thickness of the Si3N4, ITO, and Ar/CHF3 plasma contaminant should be thinner than 12.5 nm, 7.5 nm, and 100 nm, respectively. |
---|---|
ISSN: | 0034-6748 1089-7623 |
DOI: | 10.1063/5.0031869 |