Experimental set-up for Boltzmann constant determination by equipartition theorem for capacitors
A new experimental set-up for Boltzmann constant measurement is described. Statistically averaged square of voltage 〈U2〉 is measured for different capacitances C and the Boltzmann constant is determined by the equipartition theorem C 〈U2〉 = kB T. The set-up consists of low-noise high frequency opera...
Gespeichert in:
Hauptverfasser: | , , , , , |
---|---|
Format: | Tagungsbericht |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A new experimental set-up for Boltzmann constant measurement is described. Statistically averaged square of voltage 〈U2〉 is measured for different capacitances C and the Boltzmann constant is determined by the equipartition theorem C 〈U2〉 = kB T. The set-up consists of low-noise high frequency operational amplifiers ADA4898-2. An instrumental amplifier is followed by an inverting amplifier, square of the voltage is created by an analog multiplier AD633 and finally the averaged signal is measured by a multimeter. More than 10 high-school students were able to measure the Boltzmann constant with the experimental set-up in the 5th Experimental Physics Olympiad with excellent accuracy compared to the price, conditions and available time for the experiment. For the understanding of the work of the experimental set-up, it is necessary to know only the equipartition theorem. |
---|---|
ISSN: | 0094-243X 1551-7616 |
DOI: | 10.1063/1.5091404 |