An Integrating and Differentiating Bar‐Magnet Velocity Meter for Use in Ballistocardiography

An integrating and differentiating bar‐magnet velocity meter is described. Design criteria are presented for integrators and differentiators to meet given specifications of frequency response. In the design of the differentiator, use is made of a small amount of resonance magnification in a suitably...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of scientific instruments 1954-10, Vol.25 (10), p.983-988
Hauptverfasser: Perls, Thomas A., Kissinger, Charles W.
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:An integrating and differentiating bar‐magnet velocity meter is described. Design criteria are presented for integrators and differentiators to meet given specifications of frequency response. In the design of the differentiator, use is made of a small amount of resonance magnification in a suitably damped series‐resonant circuit. Test procedures are described for determinations of frequency response, sensitivity, and linear range of the instrument. Examples are given of sensitivity adjustment and installation procedures for use of this instrument in ballistocardiography. A ``60 cps hum compensator'' is used to cancel out any objectionable 60‐cycle pickup from the power line.
ISSN:0034-6748
1089-7623
DOI:10.1063/1.1770933