High temperature electrochemical scanning tunneling microscope instrument
We present a novel construction of a scanning tunneling microscope (STM) for investigations of fluid/solid interfaces and, in particular, for in situ electrochemical measurements at elevated temperatures. A special feature of this instrument is a vacuum tight connection of the electrochemical cell w...
Gespeichert in:
Veröffentlicht in: | Review of scientific instruments 2002-01, Vol.73 (1), p.102-107 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We present a novel construction of a scanning tunneling microscope (STM) for investigations of fluid/solid interfaces and, in particular, for in situ electrochemical measurements at elevated temperatures. A special feature of this instrument is a vacuum tight connection of the electrochemical cell with the STM scanner via a flexible metal bellow. This enables measurements with highly reactive and volatile fluids at high temperatures. Details of the mechanical and electronic parts of this setup are described. Test measurements on the electrodeposition of metals from molten salt electrolytes have been performed. The Ag deposition has been studied in an acidic room temperature molten salt composed of 1-butyl-3-methyl-imidazoliumchloride and
AlCl
3
up to 355 K. As a second example the Al deposition from molten
AlCl
3
–NaCl
has been tested up to 500 K. First results of these experiments are briefly presented. |
---|---|
ISSN: | 0034-6748 1089-7623 |
DOI: | 10.1063/1.1425776 |