Comparative tests of conventional and retarding-potential Mott polarimeters
The performance of a spherical field-free and a conical retarding-potential Mott polarimeter is compared. The stability of the detector signal with respect to a change in the position of the incoming electron beam is studied for two different primary electron beam energies. Shifting the incoming ele...
Gespeichert in:
Veröffentlicht in: | Review of scientific instruments 2001-09, Vol.72 (9), p.3728-3730 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The performance of a spherical field-free and a conical retarding-potential Mott polarimeter is compared. The stability of the detector signal with respect to a change in the position of the incoming electron beam is studied for two different primary electron beam energies. Shifting the incoming electron beam by 0.6 mm does not change the counting rate in the spherical field-free detector but induces a 7% or 18% change in the conical retarding-potential detector when using 1600 and 500 eV electrons, respectively. This may result in an error of the measured electron spin polarization. |
---|---|
ISSN: | 0034-6748 1089-7623 |
DOI: | 10.1063/1.1396658 |