A rotating coil probe for the magnetic field measurement on a long pulsed tokamak

In order to determine the magnitude and the position of the plasma current in a long pulsed tokamak such as the International Thermonuclear Experimental Reactor (ITER), it is urged to establish a reliable method which is free from the zero-level drift of the integrator as well as the radiation damag...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of Scientific Instruments 1999-01, Vol.70 (1), p.445-448
Hauptverfasser: Fujita, J., Matsuura, K., Kawahata, K., Fujiwaka, S., Itoh, S., Sato, K. N., Nakamura, K., Zushi, H., Sakamoto, M., Hanada, K., Jotaki, E., Makino, K.
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In order to determine the magnitude and the position of the plasma current in a long pulsed tokamak such as the International Thermonuclear Experimental Reactor (ITER), it is urged to establish a reliable method which is free from the zero-level drift of the integrator as well as the radiation damage for the steady-state magnetic field measurement. For this purpose, we have developed a hybrid system, a combination of a conventional magnetic probe for the measurement of fast varying magnetic field and a rotating coil magnetic probe for that of slowly varying field. The rotating coil is energized by an air turbine to avoid electromagnetic interference and the induce signal with a constant rotation frequency is picked up through a transformer to eliminate mechanical contacts. An automatic gain control circuit is also designed for the compensation of rotation speed fluctuation. The system is proved to achieve a flat frequency response with a proper choice of cross over frequency for high- and low-frequency systems. The rotating coil probe is tested for over 170 h without any trouble. The probe was applied to the poloidal magnetic field measurement on the TRIAM-1M long pulsed tokamak, and proved to work satisfactorily.
ISSN:0034-6748
1089-7623
DOI:10.1063/1.1149281