Battery-operated atomic force microscope
The design of a battery-operated atomic force microscope (AFM) using a piezoresistive cantilever is described. The AFM is designed so that all power to drive the scanning tube and detection electronics comes from a self-contained battery. The prototype AFM uses a 6 V, Ni–Cd, camcorder battery, howev...
Gespeichert in:
Veröffentlicht in: | Review of scientific instruments 1998-01, Vol.69 (1), p.215-220 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The design of a battery-operated atomic force microscope (AFM) using a piezoresistive cantilever is described. The AFM is designed so that all power to drive the scanning tube and detection electronics comes from a self-contained battery. The prototype AFM uses a 6 V, Ni–Cd, camcorder battery, however, any battery that supplies between 6 and 12 V may be used. Scanner control and data acquisition are implemented using commercially available software running on an external computer. The prototype AFM achieves a scan area of 53 by 53 μm, consumes 1.8 W of power, and can scan continuously for about 7 h on a single battery charge. |
---|---|
ISSN: | 0034-6748 1089-7623 |
DOI: | 10.1063/1.1148498 |