Two wire thermocouple: Frequency response in constant flow

Experimental measurements are made with a novel two wire thermocouple. Signals from two wires of unequal diameters are recorded from the thermocouple suspended in constant flow with a periodic temperature fluctuation. It is demonstrated that the reconstructed signal from the two wire thermocouple re...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of scientific instruments 1994-10, Vol.65 (10), p.3252-3257
Hauptverfasser: Forney, L. J., Fralick, G. C.
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Experimental measurements are made with a novel two wire thermocouple. Signals from two wires of unequal diameters are recorded from the thermocouple suspended in constant flow with a periodic temperature fluctuation. It is demonstrated that the reconstructed signal from the two wire thermocouple requires no compensation for ω≤2ω1 where ω1 is the natural frequency of the smaller wire. A compensation factor is recommended for larger frequencies ω≳2ω1.
ISSN:0034-6748
1089-7623
DOI:10.1063/1.1144559