Mass spectrometric studies of material evolution from magnetic liquid seals

Evolution of ferrofluid contained in two vacuum rotary feedthroughs is studied for the first time by a time‐of‐flight mass spectrometer in order to provide information on the background material these seals introduce into a high vacuum system. The seals used diester and poly(phenyl ether) based ferr...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of scientific instruments 1980-10, Vol.51 (10), p.1370-1373
Hauptverfasser: Raj, K., Grayson, M. A.
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Evolution of ferrofluid contained in two vacuum rotary feedthroughs is studied for the first time by a time‐of‐flight mass spectrometer in order to provide information on the background material these seals introduce into a high vacuum system. The seals used diester and poly(phenyl ether) based ferrofluids. The rate of evolution of the diester was found to be of the order of 3×10−10 g/s and that of the poly(phenyl ether) was below the detection limit of the mass spectrometer, i.e., 10−10 g/s. The oxygen and water evolution, which are generally of prime interest in vacuum application, were found to be 10−9 g/s for both seals at 3,000 rpm. The feedthroughs were successfully employed in precision abrasion mass spectrometric experiments.
ISSN:0034-6748
1089-7623
DOI:10.1063/1.1136067