Tonometric and Oscillometric Methods for Measurement of Central Blood Pressure Parameters: a Comparison in Patients with Borderline Hypertension or Stage 1 Hypertension

Abstract Background Changes in arterial compliance are among the first changes detectable in hypertensive syndrome. Methods with good reproducibility as compared with the gold standard for identifying such changes are desirable in clinical practice. Objectives To compare central pressure measurement...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:International journal of cardiovascular sciences 2020-04, Vol.33 (2), p.145-150
Hauptverfasser: Barroso, Weimar Kunz Sebba, Gonçalves, Claudia Ferreira, Berigó, João Alexandre Costa, Melo, Milena Andrade, Arantes, Ana Carolina, Lelis, Ellen De Souza, Sousa, Watila Moura, Rezende, Jeeziane Marcelino, Jardim, Thiago, Jardim, Paulo Cesar, Sousa, Ana Luiza Lima, Vitorino, Priscila Valverde Oliveira
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 150
container_issue 2
container_start_page 145
container_title International journal of cardiovascular sciences
container_volume 33
creator Barroso, Weimar Kunz Sebba
Gonçalves, Claudia Ferreira
Berigó, João Alexandre Costa
Melo, Milena Andrade
Arantes, Ana Carolina
Lelis, Ellen De Souza
Sousa, Watila Moura
Rezende, Jeeziane Marcelino
Jardim, Thiago
Jardim, Paulo Cesar
Sousa, Ana Luiza Lima
Vitorino, Priscila Valverde Oliveira
description Abstract Background Changes in arterial compliance are among the first changes detectable in hypertensive syndrome. Methods with good reproducibility as compared with the gold standard for identifying such changes are desirable in clinical practice. Objectives To compare central pressure measurements and arterial stiffness obtained by two non-invasive methods (tonometry and oscillometry). Methods This was a descriptive, cross-sectional study with a convenience sample of patients with borderline hypertension or stage 1 hypertension. Peripheral and central blood pressure measurements were obtained by tonometry (SphygmoCor®), considered the gold standard, and oscillometry (Mobil O´graph®). Comparisons of results were made by unpaired t-test, and p values < 0.05 were considered statistically significant. Results No difference was found in central pressure measurements obtained by SphygmoCor® (117 x 80.1 mmHg) compared with Mobil O’graph (112 x 81.4 mmHg). Mean augmentation index (AIx) was 26.1% and 21.3%, and mean pulse pressure (PP) amplification 10.7 mmHg and 10.0 mmHg by Sphygmocor® and Mobil O´graph®, respectively (p > 0.05). Mean pressure wave velocity (PWV), 8.4 m/s vs. 7.4 m/s (p = 0.013) and mean central pulse pressure, 37.7 mmmHg and 30.9 mmHg (p = 0.013) were significantly higher by SphygmoCor® than Mobil O´graph®. Conclusion Values of central systolic blood pressure, AIx and pulse pressure amplification obtained by oscillometry were not statistically different compared with tonometry; values of PWV and cPP, however, were underestimated by oscillometry. (Int J Cardiovasc Sci. 2020; 33(2):145-150)
doi_str_mv 10.36660/ijcs.20190020
format Article
fullrecord <record><control><sourceid>scielo</sourceid><recordid>TN_cdi_scielo_journals_S2359_56472020000200145</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><scielo_id>S2359_56472020000200145</scielo_id><sourcerecordid>S2359_56472020000200145</sourcerecordid><originalsourceid>FETCH-LOGICAL-c1940-ca61027c64d6e467bf73dd79c424b98cc1c2afb3482c3069ea70b63f53397e783</originalsourceid><addsrcrecordid>eNpVkN1KAzEQhYMoWLS3XucFtuZvk8Y7u6gVKi20XpdsdtambJOSbBHfyMc0iwp6NYczc-bAh9ANJRMupSS3bm_ThBGqCWHkDI0YL3VRSqHO_-hLNE5pT_IJVURLMUKfm-DDAfroLDa-wctkXdf9Oi_Q70KTcBti1iadIhzA9zi0uMozmg7PuhAavIqQhi1emWhyGGK6wwZX4XA00aXgsfN517ucSvjd9Ts8C7GB2DkPeP5xhNiDTy4f5qp1b94A03_-NbpoTZdg_DOv0Ovjw6aaF4vl03N1vygs1YIU1khKmLJSNBKEVHWreNMobQUTtZ5aSy0zbc3FlFlOpAajSC15W3KuFagpv0KT778ZBHRhuw-n6HPhdj1Q3A4UWSY8MCSEipJ_AWv3dIQ</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Tonometric and Oscillometric Methods for Measurement of Central Blood Pressure Parameters: a Comparison in Patients with Borderline Hypertension or Stage 1 Hypertension</title><source>DOAJ Directory of Open Access Journals</source><creator>Barroso, Weimar Kunz Sebba ; Gonçalves, Claudia Ferreira ; Berigó, João Alexandre Costa ; Melo, Milena Andrade ; Arantes, Ana Carolina ; Lelis, Ellen De Souza ; Sousa, Watila Moura ; Rezende, Jeeziane Marcelino ; Jardim, Thiago ; Jardim, Paulo Cesar ; Sousa, Ana Luiza Lima ; Vitorino, Priscila Valverde Oliveira</creator><creatorcontrib>Barroso, Weimar Kunz Sebba ; Gonçalves, Claudia Ferreira ; Berigó, João Alexandre Costa ; Melo, Milena Andrade ; Arantes, Ana Carolina ; Lelis, Ellen De Souza ; Sousa, Watila Moura ; Rezende, Jeeziane Marcelino ; Jardim, Thiago ; Jardim, Paulo Cesar ; Sousa, Ana Luiza Lima ; Vitorino, Priscila Valverde Oliveira</creatorcontrib><description>Abstract Background Changes in arterial compliance are among the first changes detectable in hypertensive syndrome. Methods with good reproducibility as compared with the gold standard for identifying such changes are desirable in clinical practice. Objectives To compare central pressure measurements and arterial stiffness obtained by two non-invasive methods (tonometry and oscillometry). Methods This was a descriptive, cross-sectional study with a convenience sample of patients with borderline hypertension or stage 1 hypertension. Peripheral and central blood pressure measurements were obtained by tonometry (SphygmoCor®), considered the gold standard, and oscillometry (Mobil O´graph®). Comparisons of results were made by unpaired t-test, and p values &lt; 0.05 were considered statistically significant. Results No difference was found in central pressure measurements obtained by SphygmoCor® (117 x 80.1 mmHg) compared with Mobil O’graph (112 x 81.4 mmHg). Mean augmentation index (AIx) was 26.1% and 21.3%, and mean pulse pressure (PP) amplification 10.7 mmHg and 10.0 mmHg by Sphygmocor® and Mobil O´graph®, respectively (p &gt; 0.05). Mean pressure wave velocity (PWV), 8.4 m/s vs. 7.4 m/s (p = 0.013) and mean central pulse pressure, 37.7 mmmHg and 30.9 mmHg (p = 0.013) were significantly higher by SphygmoCor® than Mobil O´graph®. Conclusion Values of central systolic blood pressure, AIx and pulse pressure amplification obtained by oscillometry were not statistically different compared with tonometry; values of PWV and cPP, however, were underestimated by oscillometry. (Int J Cardiovasc Sci. 2020; 33(2):145-150)</description><identifier>ISSN: 2359-5647</identifier><identifier>EISSN: 2359-5647</identifier><identifier>DOI: 10.36660/ijcs.20190020</identifier><language>eng</language><publisher>Sociedade Brasileira de Cardiologia</publisher><subject>CARDIAC &amp; CARDIOVASCULAR SYSTEMS</subject><ispartof>International journal of cardiovascular sciences, 2020-04, Vol.33 (2), p.145-150</ispartof><rights>This work is licensed under a Creative Commons Attribution 4.0 International License.</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c1940-ca61027c64d6e467bf73dd79c424b98cc1c2afb3482c3069ea70b63f53397e783</citedby></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>230,314,780,784,864,885,27923,27924</link.rule.ids></links><search><creatorcontrib>Barroso, Weimar Kunz Sebba</creatorcontrib><creatorcontrib>Gonçalves, Claudia Ferreira</creatorcontrib><creatorcontrib>Berigó, João Alexandre Costa</creatorcontrib><creatorcontrib>Melo, Milena Andrade</creatorcontrib><creatorcontrib>Arantes, Ana Carolina</creatorcontrib><creatorcontrib>Lelis, Ellen De Souza</creatorcontrib><creatorcontrib>Sousa, Watila Moura</creatorcontrib><creatorcontrib>Rezende, Jeeziane Marcelino</creatorcontrib><creatorcontrib>Jardim, Thiago</creatorcontrib><creatorcontrib>Jardim, Paulo Cesar</creatorcontrib><creatorcontrib>Sousa, Ana Luiza Lima</creatorcontrib><creatorcontrib>Vitorino, Priscila Valverde Oliveira</creatorcontrib><title>Tonometric and Oscillometric Methods for Measurement of Central Blood Pressure Parameters: a Comparison in Patients with Borderline Hypertension or Stage 1 Hypertension</title><title>International journal of cardiovascular sciences</title><addtitle>Int. J. Cardiovasc. Sci</addtitle><description>Abstract Background Changes in arterial compliance are among the first changes detectable in hypertensive syndrome. Methods with good reproducibility as compared with the gold standard for identifying such changes are desirable in clinical practice. Objectives To compare central pressure measurements and arterial stiffness obtained by two non-invasive methods (tonometry and oscillometry). Methods This was a descriptive, cross-sectional study with a convenience sample of patients with borderline hypertension or stage 1 hypertension. Peripheral and central blood pressure measurements were obtained by tonometry (SphygmoCor®), considered the gold standard, and oscillometry (Mobil O´graph®). Comparisons of results were made by unpaired t-test, and p values &lt; 0.05 were considered statistically significant. Results No difference was found in central pressure measurements obtained by SphygmoCor® (117 x 80.1 mmHg) compared with Mobil O’graph (112 x 81.4 mmHg). Mean augmentation index (AIx) was 26.1% and 21.3%, and mean pulse pressure (PP) amplification 10.7 mmHg and 10.0 mmHg by Sphygmocor® and Mobil O´graph®, respectively (p &gt; 0.05). Mean pressure wave velocity (PWV), 8.4 m/s vs. 7.4 m/s (p = 0.013) and mean central pulse pressure, 37.7 mmmHg and 30.9 mmHg (p = 0.013) were significantly higher by SphygmoCor® than Mobil O´graph®. Conclusion Values of central systolic blood pressure, AIx and pulse pressure amplification obtained by oscillometry were not statistically different compared with tonometry; values of PWV and cPP, however, were underestimated by oscillometry. (Int J Cardiovasc Sci. 2020; 33(2):145-150)</description><subject>CARDIAC &amp; CARDIOVASCULAR SYSTEMS</subject><issn>2359-5647</issn><issn>2359-5647</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2020</creationdate><recordtype>article</recordtype><recordid>eNpVkN1KAzEQhYMoWLS3XucFtuZvk8Y7u6gVKi20XpdsdtambJOSbBHfyMc0iwp6NYczc-bAh9ANJRMupSS3bm_ThBGqCWHkDI0YL3VRSqHO_-hLNE5pT_IJVURLMUKfm-DDAfroLDa-wctkXdf9Oi_Q70KTcBti1iadIhzA9zi0uMozmg7PuhAavIqQhi1emWhyGGK6wwZX4XA00aXgsfN517ucSvjd9Ts8C7GB2DkPeP5xhNiDTy4f5qp1b94A03_-NbpoTZdg_DOv0Ovjw6aaF4vl03N1vygs1YIU1khKmLJSNBKEVHWreNMobQUTtZ5aSy0zbc3FlFlOpAajSC15W3KuFagpv0KT778ZBHRhuw-n6HPhdj1Q3A4UWSY8MCSEipJ_AWv3dIQ</recordid><startdate>20200401</startdate><enddate>20200401</enddate><creator>Barroso, Weimar Kunz Sebba</creator><creator>Gonçalves, Claudia Ferreira</creator><creator>Berigó, João Alexandre Costa</creator><creator>Melo, Milena Andrade</creator><creator>Arantes, Ana Carolina</creator><creator>Lelis, Ellen De Souza</creator><creator>Sousa, Watila Moura</creator><creator>Rezende, Jeeziane Marcelino</creator><creator>Jardim, Thiago</creator><creator>Jardim, Paulo Cesar</creator><creator>Sousa, Ana Luiza Lima</creator><creator>Vitorino, Priscila Valverde Oliveira</creator><general>Sociedade Brasileira de Cardiologia</general><scope>GPN</scope></search><sort><creationdate>20200401</creationdate><title>Tonometric and Oscillometric Methods for Measurement of Central Blood Pressure Parameters: a Comparison in Patients with Borderline Hypertension or Stage 1 Hypertension</title><author>Barroso, Weimar Kunz Sebba ; Gonçalves, Claudia Ferreira ; Berigó, João Alexandre Costa ; Melo, Milena Andrade ; Arantes, Ana Carolina ; Lelis, Ellen De Souza ; Sousa, Watila Moura ; Rezende, Jeeziane Marcelino ; Jardim, Thiago ; Jardim, Paulo Cesar ; Sousa, Ana Luiza Lima ; Vitorino, Priscila Valverde Oliveira</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c1940-ca61027c64d6e467bf73dd79c424b98cc1c2afb3482c3069ea70b63f53397e783</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2020</creationdate><topic>CARDIAC &amp; CARDIOVASCULAR SYSTEMS</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Barroso, Weimar Kunz Sebba</creatorcontrib><creatorcontrib>Gonçalves, Claudia Ferreira</creatorcontrib><creatorcontrib>Berigó, João Alexandre Costa</creatorcontrib><creatorcontrib>Melo, Milena Andrade</creatorcontrib><creatorcontrib>Arantes, Ana Carolina</creatorcontrib><creatorcontrib>Lelis, Ellen De Souza</creatorcontrib><creatorcontrib>Sousa, Watila Moura</creatorcontrib><creatorcontrib>Rezende, Jeeziane Marcelino</creatorcontrib><creatorcontrib>Jardim, Thiago</creatorcontrib><creatorcontrib>Jardim, Paulo Cesar</creatorcontrib><creatorcontrib>Sousa, Ana Luiza Lima</creatorcontrib><creatorcontrib>Vitorino, Priscila Valverde Oliveira</creatorcontrib><collection>SciELO</collection><jtitle>International journal of cardiovascular sciences</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Barroso, Weimar Kunz Sebba</au><au>Gonçalves, Claudia Ferreira</au><au>Berigó, João Alexandre Costa</au><au>Melo, Milena Andrade</au><au>Arantes, Ana Carolina</au><au>Lelis, Ellen De Souza</au><au>Sousa, Watila Moura</au><au>Rezende, Jeeziane Marcelino</au><au>Jardim, Thiago</au><au>Jardim, Paulo Cesar</au><au>Sousa, Ana Luiza Lima</au><au>Vitorino, Priscila Valverde Oliveira</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Tonometric and Oscillometric Methods for Measurement of Central Blood Pressure Parameters: a Comparison in Patients with Borderline Hypertension or Stage 1 Hypertension</atitle><jtitle>International journal of cardiovascular sciences</jtitle><addtitle>Int. J. Cardiovasc. Sci</addtitle><date>2020-04-01</date><risdate>2020</risdate><volume>33</volume><issue>2</issue><spage>145</spage><epage>150</epage><pages>145-150</pages><issn>2359-5647</issn><eissn>2359-5647</eissn><abstract>Abstract Background Changes in arterial compliance are among the first changes detectable in hypertensive syndrome. Methods with good reproducibility as compared with the gold standard for identifying such changes are desirable in clinical practice. Objectives To compare central pressure measurements and arterial stiffness obtained by two non-invasive methods (tonometry and oscillometry). Methods This was a descriptive, cross-sectional study with a convenience sample of patients with borderline hypertension or stage 1 hypertension. Peripheral and central blood pressure measurements were obtained by tonometry (SphygmoCor®), considered the gold standard, and oscillometry (Mobil O´graph®). Comparisons of results were made by unpaired t-test, and p values &lt; 0.05 were considered statistically significant. Results No difference was found in central pressure measurements obtained by SphygmoCor® (117 x 80.1 mmHg) compared with Mobil O’graph (112 x 81.4 mmHg). Mean augmentation index (AIx) was 26.1% and 21.3%, and mean pulse pressure (PP) amplification 10.7 mmHg and 10.0 mmHg by Sphygmocor® and Mobil O´graph®, respectively (p &gt; 0.05). Mean pressure wave velocity (PWV), 8.4 m/s vs. 7.4 m/s (p = 0.013) and mean central pulse pressure, 37.7 mmmHg and 30.9 mmHg (p = 0.013) were significantly higher by SphygmoCor® than Mobil O´graph®. Conclusion Values of central systolic blood pressure, AIx and pulse pressure amplification obtained by oscillometry were not statistically different compared with tonometry; values of PWV and cPP, however, were underestimated by oscillometry. (Int J Cardiovasc Sci. 2020; 33(2):145-150)</abstract><pub>Sociedade Brasileira de Cardiologia</pub><doi>10.36660/ijcs.20190020</doi><tpages>6</tpages><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 2359-5647
ispartof International journal of cardiovascular sciences, 2020-04, Vol.33 (2), p.145-150
issn 2359-5647
2359-5647
language eng
recordid cdi_scielo_journals_S2359_56472020000200145
source DOAJ Directory of Open Access Journals
subjects CARDIAC & CARDIOVASCULAR SYSTEMS
title Tonometric and Oscillometric Methods for Measurement of Central Blood Pressure Parameters: a Comparison in Patients with Borderline Hypertension or Stage 1 Hypertension
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-11T21%3A46%3A32IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-scielo&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Tonometric%20and%20Oscillometric%20Methods%20for%20Measurement%20of%20Central%20Blood%20Pressure%20Parameters:%20a%20Comparison%20in%20Patients%20with%20Borderline%20Hypertension%20or%20Stage%201%20Hypertension&rft.jtitle=International%20journal%20of%20cardiovascular%20sciences&rft.au=Barroso,%20Weimar%20Kunz%20Sebba&rft.date=2020-04-01&rft.volume=33&rft.issue=2&rft.spage=145&rft.epage=150&rft.pages=145-150&rft.issn=2359-5647&rft.eissn=2359-5647&rft_id=info:doi/10.36660/ijcs.20190020&rft_dat=%3Cscielo%3ES2359_56472020000200145%3C/scielo%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_id=info:pmid/&rft_scielo_id=S2359_56472020000200145&rfr_iscdi=true