Tonometric and Oscillometric Methods for Measurement of Central Blood Pressure Parameters: a Comparison in Patients with Borderline Hypertension or Stage 1 Hypertension

Abstract Background Changes in arterial compliance are among the first changes detectable in hypertensive syndrome. Methods with good reproducibility as compared with the gold standard for identifying such changes are desirable in clinical practice. Objectives To compare central pressure measurement...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:International journal of cardiovascular sciences 2020-04, Vol.33 (2), p.145-150
Hauptverfasser: Barroso, Weimar Kunz Sebba, Gonçalves, Claudia Ferreira, Berigó, João Alexandre Costa, Melo, Milena Andrade, Arantes, Ana Carolina, Lelis, Ellen De Souza, Sousa, Watila Moura, Rezende, Jeeziane Marcelino, Jardim, Thiago, Jardim, Paulo Cesar, Sousa, Ana Luiza Lima, Vitorino, Priscila Valverde Oliveira
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract Background Changes in arterial compliance are among the first changes detectable in hypertensive syndrome. Methods with good reproducibility as compared with the gold standard for identifying such changes are desirable in clinical practice. Objectives To compare central pressure measurements and arterial stiffness obtained by two non-invasive methods (tonometry and oscillometry). Methods This was a descriptive, cross-sectional study with a convenience sample of patients with borderline hypertension or stage 1 hypertension. Peripheral and central blood pressure measurements were obtained by tonometry (SphygmoCor®), considered the gold standard, and oscillometry (Mobil O´graph®). Comparisons of results were made by unpaired t-test, and p values < 0.05 were considered statistically significant. Results No difference was found in central pressure measurements obtained by SphygmoCor® (117 x 80.1 mmHg) compared with Mobil O’graph (112 x 81.4 mmHg). Mean augmentation index (AIx) was 26.1% and 21.3%, and mean pulse pressure (PP) amplification 10.7 mmHg and 10.0 mmHg by Sphygmocor® and Mobil O´graph®, respectively (p > 0.05). Mean pressure wave velocity (PWV), 8.4 m/s vs. 7.4 m/s (p = 0.013) and mean central pulse pressure, 37.7 mmmHg and 30.9 mmHg (p = 0.013) were significantly higher by SphygmoCor® than Mobil O´graph®. Conclusion Values of central systolic blood pressure, AIx and pulse pressure amplification obtained by oscillometry were not statistically different compared with tonometry; values of PWV and cPP, however, were underestimated by oscillometry. (Int J Cardiovasc Sci. 2020; 33(2):145-150)
ISSN:2359-5647
2359-5647
DOI:10.36660/ijcs.20190020