Wavelet-Based determination of Cu and Pb in water samples using potentiometric stripping analysis with a Lab-Made potentiostat

This paper demonstrates the possibility of simultaneously determining lead and copper in mineral, river and tap water by potentiometric stripping analysis using a low-cost lab-made potentiostat and a wavelet-based signal processing method. Pb(II) and Cu(II) concentrations were found to be in the ran...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of the Brazilian Chemical Society 2009, Vol.20 (8), p.1561-1564
Hauptverfasser: Martins, Valdomiro L., Galvão, Roberto K. Harrop, Araújo, Mário César U. de, Gaião, Edvaldo da Nóbrega
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This paper demonstrates the possibility of simultaneously determining lead and copper in mineral, river and tap water by potentiometric stripping analysis using a low-cost lab-made potentiostat and a wavelet-based signal processing method. Pb(II) and Cu(II) concentrations were found to be in the range 3-7 µg L-1. Relative standard deviations were on average smaller than 3%, which is satisfactory in view of the trace concentration. For comparison, the determination was also carried out by using a high-end commercial equipment (μAutolab Type II, Eco Chemie). The average and standard deviation of the results were similar to those obtained with the lab-made potentiostat. No significant statistical difference was verified according to a paired t-test at a confidence level of 95%.
ISSN:0103-5053
1678-4790