Impact of chamber pressure and material properties on the deformation response of corneal models measured by dynamic ultra-high-speed Scheimpflug imaging

To study the deformation response of three distinct contact lenses with known structures, which served as corneal models, under different chamber pressures using ultra-high-speed (UHS) Scheimpflug imaging. Three hydrophilic contact lenses were mounted on a sealed water chamber with precisely adjusta...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Arquivos brasileiros de oftalmologia 2013-10, Vol.76 (5), p.278-281
Hauptverfasser: Correia, Fernando Faria, Ramos, Isaac, Roberts, Cynthia J, Steinmueller, Andreas, Krug, Matthias, Ambrósio, Jr, Renato
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:To study the deformation response of three distinct contact lenses with known structures, which served as corneal models, under different chamber pressures using ultra-high-speed (UHS) Scheimpflug imaging. Three hydrophilic contact lenses were mounted on a sealed water chamber with precisely adjustable pressure: TAN-G5X (41% hydroxyethylmethacrylate/glycolmethacrylate, 550 µm thick), TAN-40 (62% hydroxyethylmethacrylate, 525 µm thick) and TAN-58 (42% methylmethacrylate, 258 µm thick). Each model was tested five times under different pressures (5, 15, 25, 35 and 45 mmHg), using ultra-high-speed Scheimpflug imaging during non-contact tonometry. 140 Scheimpflug images were taken with the UHS camera in each measurement. The deformation amplitude during non-contact tonometry was determined as the highest displacement of the apex at the highest concavity (HC) moment. At each pressure level, the deformation amplitude was statistically different for each lens tested (p
ISSN:0004-2749
1678-2925
1678-2925
DOI:10.1590/S0004-27492013000500005