Comparison between a newly developed PC-based Doppler umbilical artery waveform analyser and a commercial unit : original article

Objectives. To determine the accuracy of the resistance index (RI) of flow velocity waveforms of the umbilical artery measured using a newly developed PC-based continuous wave Doppler device (Umbiflow) with regard to systematic and random variations when compared with a commercial standard (Vasoflow...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:South African medical journal 2005-01, Vol.95 (1), p.62-64
Hauptverfasser: Bunn, A.E., Theron, G.B., Odendaal, H.J., Theron, A.M.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 64
container_issue 1
container_start_page 62
container_title South African medical journal
container_volume 95
creator Bunn, A.E.
Theron, G.B.
Odendaal, H.J.
Theron, A.M.
description Objectives. To determine the accuracy of the resistance index (RI) of flow velocity waveforms of the umbilical artery measured using a newly developed PC-based continuous wave Doppler device (Umbiflow) with regard to systematic and random variations when compared with a commercial standard (Vasoflow). Design. A cohort study. Setting. The fetal evaluation clinic (FEC) at Tygerberg Hospital. Subjects. Patients referred to the FEC at Tygerberg Hospital with suspected chronic placental insufficiency. Outcome measures. The correlation coefficients indicating the strength of the relationship between the two devices and their agreement using the method of Bland and Altman. Results. A total of 248 patients were studied. The mean RIs of the first Doppler assessment were 0.69 (standard deviation (SD) 0.11) and 0.67 (SD 0.11) using the Vasoflow and Umbiflow respectively. The Pearson's correlation coefficient comparing the RI of the first test was 0.85. The degree of agreement between the two methods was excellent, the mean differences being very small (Conclusions. The accuracy of the Umbiflow has been proved. A non-significant trend towards slightly lower values needs to be considered. This could be addressed by changing the percentile cut-off to slightly lower values.
format Article
fullrecord <record><control><sourceid>sabinet_JRA</sourceid><recordid>TN_cdi_sabinet_saepub_10520_EJC68320</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sabinet_id>10520/EJC68320</sabinet_id><sourcerecordid>10520/EJC68320</sourcerecordid><originalsourceid>FETCH-sabinet_saepub_10520_EJC683203</originalsourceid><addsrcrecordid>eNqNjMtqwzAQRUVoIO7jH2aRrUGxK-exdVNKV110b0b2JEzRC8mO8bJ_XkH7AV3dc-HcuxJFJfeHUu1qdScKWammPKr980bcp_Qlc1fHphDfrbcBIyfvQNM4EzlAcDSbBQa6kfGBBvhoS40pw4sPwVCEyWo23KMBjCPFBWa80cVHC-jQLCkr6Ib81HtrKfaczcnxCCfwka_sfpfcG3oU6wuaRE9_-SC2r-fP9q1MqNnR2CWkMOluJ1Ulu_N72xzqStb_1H4AX9BRzg</addsrcrecordid><sourcetype>Publisher</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Comparison between a newly developed PC-based Doppler umbilical artery waveform analyser and a commercial unit : original article</title><source>Sabinet Open Access Journals</source><creator>Bunn, A.E. ; Theron, G.B. ; Odendaal, H.J. ; Theron, A.M.</creator><creatorcontrib>Bunn, A.E. ; Theron, G.B. ; Odendaal, H.J. ; Theron, A.M.</creatorcontrib><description>Objectives. To determine the accuracy of the resistance index (RI) of flow velocity waveforms of the umbilical artery measured using a newly developed PC-based continuous wave Doppler device (Umbiflow) with regard to systematic and random variations when compared with a commercial standard (Vasoflow). Design. A cohort study. Setting. The fetal evaluation clinic (FEC) at Tygerberg Hospital. Subjects. Patients referred to the FEC at Tygerberg Hospital with suspected chronic placental insufficiency. Outcome measures. The correlation coefficients indicating the strength of the relationship between the two devices and their agreement using the method of Bland and Altman. Results. A total of 248 patients were studied. The mean RIs of the first Doppler assessment were 0.69 (standard deviation (SD) 0.11) and 0.67 (SD 0.11) using the Vasoflow and Umbiflow respectively. The Pearson's correlation coefficient comparing the RI of the first test was 0.85. The degree of agreement between the two methods was excellent, the mean differences being very small (Conclusions. The accuracy of the Umbiflow has been proved. A non-significant trend towards slightly lower values needs to be considered. This could be addressed by changing the percentile cut-off to slightly lower values.</description><identifier>ISSN: 0256-9574</identifier><identifier>EISSN: 2078-5135</identifier><language>eng</language><publisher>Health and Medical Publishing Group (HMPG)</publisher><subject>Comparative studies ; Flowvelocity waveforms ; Resistance index ; Umbiflow ; Umbilical artery ; Vasoflow</subject><ispartof>South African medical journal, 2005-01, Vol.95 (1), p.62-64</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,776,780,39219</link.rule.ids><linktorsrc>$$Uhttp://hdl.handle.net/10520/EJC68320$$EView_record_in_Sabinet_Online_Ltd.$$FView_record_in_$$GSabinet_Online_Ltd.</linktorsrc></links><search><creatorcontrib>Bunn, A.E.</creatorcontrib><creatorcontrib>Theron, G.B.</creatorcontrib><creatorcontrib>Odendaal, H.J.</creatorcontrib><creatorcontrib>Theron, A.M.</creatorcontrib><title>Comparison between a newly developed PC-based Doppler umbilical artery waveform analyser and a commercial unit : original article</title><title>South African medical journal</title><description>Objectives. To determine the accuracy of the resistance index (RI) of flow velocity waveforms of the umbilical artery measured using a newly developed PC-based continuous wave Doppler device (Umbiflow) with regard to systematic and random variations when compared with a commercial standard (Vasoflow). Design. A cohort study. Setting. The fetal evaluation clinic (FEC) at Tygerberg Hospital. Subjects. Patients referred to the FEC at Tygerberg Hospital with suspected chronic placental insufficiency. Outcome measures. The correlation coefficients indicating the strength of the relationship between the two devices and their agreement using the method of Bland and Altman. Results. A total of 248 patients were studied. The mean RIs of the first Doppler assessment were 0.69 (standard deviation (SD) 0.11) and 0.67 (SD 0.11) using the Vasoflow and Umbiflow respectively. The Pearson's correlation coefficient comparing the RI of the first test was 0.85. The degree of agreement between the two methods was excellent, the mean differences being very small (Conclusions. The accuracy of the Umbiflow has been proved. A non-significant trend towards slightly lower values needs to be considered. This could be addressed by changing the percentile cut-off to slightly lower values.</description><subject>Comparative studies</subject><subject>Flowvelocity waveforms</subject><subject>Resistance index</subject><subject>Umbiflow</subject><subject>Umbilical artery</subject><subject>Vasoflow</subject><issn>0256-9574</issn><issn>2078-5135</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2005</creationdate><recordtype>article</recordtype><sourceid/><recordid>eNqNjMtqwzAQRUVoIO7jH2aRrUGxK-exdVNKV110b0b2JEzRC8mO8bJ_XkH7AV3dc-HcuxJFJfeHUu1qdScKWammPKr980bcp_Qlc1fHphDfrbcBIyfvQNM4EzlAcDSbBQa6kfGBBvhoS40pw4sPwVCEyWo23KMBjCPFBWa80cVHC-jQLCkr6Ib81HtrKfaczcnxCCfwka_sfpfcG3oU6wuaRE9_-SC2r-fP9q1MqNnR2CWkMOluJ1Ulu_N72xzqStb_1H4AX9BRzg</recordid><startdate>20050101</startdate><enddate>20050101</enddate><creator>Bunn, A.E.</creator><creator>Theron, G.B.</creator><creator>Odendaal, H.J.</creator><creator>Theron, A.M.</creator><general>Health and Medical Publishing Group (HMPG)</general><scope/></search><sort><creationdate>20050101</creationdate><title>Comparison between a newly developed PC-based Doppler umbilical artery waveform analyser and a commercial unit : original article</title><author>Bunn, A.E. ; Theron, G.B. ; Odendaal, H.J. ; Theron, A.M.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-sabinet_saepub_10520_EJC683203</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2005</creationdate><topic>Comparative studies</topic><topic>Flowvelocity waveforms</topic><topic>Resistance index</topic><topic>Umbiflow</topic><topic>Umbilical artery</topic><topic>Vasoflow</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Bunn, A.E.</creatorcontrib><creatorcontrib>Theron, G.B.</creatorcontrib><creatorcontrib>Odendaal, H.J.</creatorcontrib><creatorcontrib>Theron, A.M.</creatorcontrib><jtitle>South African medical journal</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext_linktorsrc</fulltext></delivery><addata><au>Bunn, A.E.</au><au>Theron, G.B.</au><au>Odendaal, H.J.</au><au>Theron, A.M.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Comparison between a newly developed PC-based Doppler umbilical artery waveform analyser and a commercial unit : original article</atitle><jtitle>South African medical journal</jtitle><date>2005-01-01</date><risdate>2005</risdate><volume>95</volume><issue>1</issue><spage>62</spage><epage>64</epage><pages>62-64</pages><issn>0256-9574</issn><eissn>2078-5135</eissn><abstract>Objectives. To determine the accuracy of the resistance index (RI) of flow velocity waveforms of the umbilical artery measured using a newly developed PC-based continuous wave Doppler device (Umbiflow) with regard to systematic and random variations when compared with a commercial standard (Vasoflow). Design. A cohort study. Setting. The fetal evaluation clinic (FEC) at Tygerberg Hospital. Subjects. Patients referred to the FEC at Tygerberg Hospital with suspected chronic placental insufficiency. Outcome measures. The correlation coefficients indicating the strength of the relationship between the two devices and their agreement using the method of Bland and Altman. Results. A total of 248 patients were studied. The mean RIs of the first Doppler assessment were 0.69 (standard deviation (SD) 0.11) and 0.67 (SD 0.11) using the Vasoflow and Umbiflow respectively. The Pearson's correlation coefficient comparing the RI of the first test was 0.85. The degree of agreement between the two methods was excellent, the mean differences being very small (Conclusions. The accuracy of the Umbiflow has been proved. A non-significant trend towards slightly lower values needs to be considered. This could be addressed by changing the percentile cut-off to slightly lower values.</abstract><pub>Health and Medical Publishing Group (HMPG)</pub></addata></record>
fulltext fulltext_linktorsrc
identifier ISSN: 0256-9574
ispartof South African medical journal, 2005-01, Vol.95 (1), p.62-64
issn 0256-9574
2078-5135
language eng
recordid cdi_sabinet_saepub_10520_EJC68320
source Sabinet Open Access Journals
subjects Comparative studies
Flowvelocity waveforms
Resistance index
Umbiflow
Umbilical artery
Vasoflow
title Comparison between a newly developed PC-based Doppler umbilical artery waveform analyser and a commercial unit : original article
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-07T19%3A38%3A17IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-sabinet_JRA&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Comparison%20between%20a%20newly%20developed%20PC-based%20Doppler%20umbilical%20artery%20waveform%20analyser%20and%20a%20commercial%20unit%20:%20original%20article&rft.jtitle=South%20African%20medical%20journal&rft.au=Bunn,%20A.E.&rft.date=2005-01-01&rft.volume=95&rft.issue=1&rft.spage=62&rft.epage=64&rft.pages=62-64&rft.issn=0256-9574&rft.eissn=2078-5135&rft_id=info:doi/&rft_dat=%3Csabinet_JRA%3E10520/EJC68320%3C/sabinet_JRA%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_id=info:pmid/&rft_sabinet_id=10520/EJC68320&rfr_iscdi=true